Cerebral oxygenation after birth – a comparison of INVOS ® and FORE ‐ SIGHT ™ near‐infrared spectroscopy oximeters
Open Access
- 20 March 2014
- journal article
- research article
- Published by Wiley in Acta Paediatrica
- Vol. 103 (5) , 488-493
Abstract
To compare absolute values of regional cerebral tissue oxygenation (cStO2 ) during haemodynamic transition after birth and repeatability during steady state for two commercial near-infrared spectroscopy (NIRS) devices. In a prospective observational study, the INVOS 5100C and FORE-SIGHT were compared on 12 term newborns delivered by elective caesarean section. During the 10 min following umbilical cord clamping, cStO2 was measured simultaneously with the neonatal sensors from each device. Repeated measurements were taken the following day. Three and 8 min after clamping, the mean cStO2 value increased from 53.4% (CI 36.8-69.9%) to 86.0% (CI 80.2-91.7%) for INVOS and from 61.6% (CI 55.4-67.8%) to 82.2% (CI 77.7-86.7%) for FORE-SIGHT. The Bland-Altman plot revealed decreasing difference (INVOS minus FORE-SIGHT) (D) in absolute values (A) with increasing cStO2 (D = 0.5A - 38.19 p = <0.001). The mean steady-state value on day two was 78.4% (CI 74.6-82.2%) and 86.2% (CI 85.0-87.4%) for INVOS and FORE-SIGHT, respectively. The within-subject standard deviation during steady-state repeated measurements was 4.8% ± 0.86 for INVOS and 2.8% ± 0.5 for FORE-SIGHT. The INVOS and FORE-SIGHT cStO2 estimates showed oxygenation-level-dependent difference during birth transition. The better repeatability of FORE-SIGHT could be due to the lower response to change in saturation.Keywords
Funding Information
- The Augustinus Foundation
This publication has 25 references indexed in Scilit:
- Peripheral tissue oximetry: comparing three commercial near-infrared spectroscopy oximeters on the forearmJournal of Clinical Monitoring and Computing, 2013
- Cerebral Near-Infrared Spectroscopy during Transition of Healthy Term NewbornsNeonatology, 2013
- Has the time come to use near-infrared spectroscopy as a routine clinical tool in preterm infants undergoing intensive care?Philosophical Transactions of the Royal Society A: Mathematical, Physical and Engineering Sciences, 2011
- Tissue oximetry: a comparison of mean values of regional tissue saturation, reproducibility and dynamic range of four NIRS-instruments on the human forearmBiomedical Optics Express, 2011
- Brain oxygenation monitoring during neonatal resuscitation of very low birth weight infantsJournal of Perinatology, 2011
- Regional tissue oxygen saturation: comparability and reproducibility of different devicesJournal of Biomedical Optics, 2011
- Regional Oxygen Saturation of the Brain and Peripheral Tissue during Birth Transition of Term InfantsThe Journal of Pediatrics, 2010
- Is near-infrared spectroscopy living up to its promises?Seminars in Fetal and Neonatal Medicine, 2006
- Reproducibility of cerebral oxygenation measurement in neonates and infants in the clinical setting using the NIRO 300 oximeter*Pediatric Critical Care Medicine, 2005
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986