Accuracy of transcutaneous bilirubin measurement in preterm low-birth-weight neonates

Authors:

Abstract
The objective of this study was to evaluate the correlation and agreement between transcutaneous and serum bilirubin among preterm low-birth-weight neonates. Neonates born at r = 0.84, mean difference, 0.3 ± 1.9 mg/dL) followed by TcB on abdomen (r = 0.73, mean difference, 1.5 ± 2.6 mg/dL) and sternum (r = 0.72, mean difference, 1.5 ± 2.6 mg/dL). TcB performed well at all three points of measurement with best correlations being observed at icterus level 4/5. Correlation between TcB and STB-H measured by high-performance liquid chromatography was less strong but significant (r = 0.59 to 0.69 at different time points of measurement). Conclusions: TcB has good correlation and agreement with STB in preterm low-birth-weight neonates born at ≥28 weeks of gestation.