Radiation Dose Savings for Adult Pulmonary Embolus 64-MDCT Using Bismuth Breast Shields, Lower Peak Kilovoltage, and Automatic Tube Current Modulation
- 1 January 2009
- journal article
- Published by American Roentgen Ray Society in American Journal of Roentgenology
- Vol. 192 (1) , 244-253
Abstract
OBJECTIVE. The purpose of this study was to assess whether radiation dose savings using a lower peak kilovoltage (kVp) setting, bismuth breast shields, and automatic tube current modulation could be achieved while preserving the image quality of MDCT scans obtained to assess for pulmonary embolus (PE).MATERIALS AND METHODS. CT angiography (CTA) examinations were performed to assess for the presence or absence of pulmonary artery emboli using a 64-MDCT scanner with automatic tube current modulation (noise level = 10 HU), two kVp settings (120 and 140 kVp), and bismuth breast shields. Absorbed organ doses were measured using anthropomorphic phantoms and metal oxide semiconductor field effect transistor (MOSFET) detectors. Image quality was assessed quantitatively as well as qualitatively in various anatomic sites of the thorax.RESULTS. Using a lower kVp (120 vs 140 kVp) and automatic tube current modulation resulted in a dose savings of 27% to the breast and 47% to the lungs. The use of a lower kVp (120 kVp...Keywords
This publication has 18 references indexed in Scilit:
- Radiation Dose from Contemporary Cardiothoracic Multidetector CT Protocols with an Anthropomorphic Female Phantom: Implications for Cancer InductionRadiology, 2007
- Image Quality and Radiation Exposure at Pulmonary CT Angiography with 100- or 120-kVp Protocol: Prospective Randomized StudyRadiology, 2007
- Effective Dose Determination Using an Anthropomorphic Phantom and Metal Oxide Semiconductor Field Effect Transistor Technology for Clinical Adult Body Multidetector Array Computed Tomography ProtocolsJournal of Computer Assisted Tomography, 2007
- Female Breast Radiation Exposure During Thorax Multidetector Computed Tomography and the Effectiveness of Bismuth Breast Shield to Reduce Breast Radiation DoseJournal of Computer Assisted Tomography, 2007
- CT Angiography of Pulmonary Arteries to Detect Pulmonary Embolism: Improvement of Vascular Enhancement with Low Kilovoltage SettingsRadiology, 2006
- Lower Tube Voltage Reduces Contrast Material and Radiation Doses on 16-MDCT AortographyAmerican Journal of Roentgenology, 2006
- MDCT of the Coronary Arteries: Feasibility of Low-Dose CT with ECG-Pulsed Tube Current Modulation to Reduce Radiation DoseAmerican Journal of Roentgenology, 2006
- Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image qualityEuropean Radiology, 2006
- Techniques and Applications of Automatic Tube Current Modulation for CTRadiology, 2004
- Low-Kilovoltage Multi–Detector Row Chest CT in Adults: Feasibility and Effect on Image Quality and Iodine DoseRadiology, 2004