Radiation Dose Savings for Adult Pulmonary Embolus 64-MDCT Using Bismuth Breast Shields, Lower Peak Kilovoltage, and Automatic Tube Current Modulation

Authors:

Abstract
OBJECTIVE. The purpose of this study was to assess whether radiation dose savings using a lower peak kilovoltage (kVp) setting, bismuth breast shields, and automatic tube current modulation could be achieved while preserving the image quality of MDCT scans obtained to assess for pulmonary embolus (PE).MATERIALS AND METHODS. CT angiography (CTA) examinations were performed to assess for the presence or absence of pulmonary artery emboli using a 64-MDCT scanner with automatic tube current modulation (noise level = 10 HU), two kVp settings (120 and 140 kVp), and bismuth breast shields. Absorbed organ doses were measured using anthropomorphic phantoms and metal oxide semiconductor field effect transistor (MOSFET) detectors. Image quality was assessed quantitatively as well as qualitatively in various anatomic sites of the thorax.RESULTS. Using a lower kVp (120 vs 140 kVp) and automatic tube current modulation resulted in a dose savings of 27% to the breast and 47% to the lungs. The use of a lower kVp (120 kVp...