Conductive Impulse for Wireless Communication in Dual-Chamber Leadless Pacemakers
- 7 December 2020
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Microwave Theory and Techniques
- Vol. 69 (1) , 443-451
Abstract
Conductive impulse signaling is proposed for high-efficiency low data-rate wireless communication for future development in cardiac leadless pacemaker technology. The conductivity of the heart muscles and blood is used for electrical signal transmission between two pacemaker implants in the right atrium and the right ventricle. The communication channel is characterized in the frequency range of 1-15 MHz using numerical electromagnetic computations. The wideband implant electrodes support the impulse signaling, and the frequency flat communication channel prevents impulse distortion that simplifies the receiver design using a threshold detector scheme. An ultralow-power impulse generator is designed using off-the-shelf electronics and is integrated into a leadless pacemaker mockup, and the impulse transmission between two capsules is measured in a homogeneous phantom and in vivo animal heart experiments. The feasibility of system integration with the state-of-the-art leadless cardiac pacemakers is demonstrated that can support ultralow-power, low data-rate wireless communications.Keywords
Funding Information
- Research Council of Norway Projects “Wireless In-Body Sensor and Actuator Networks (WINNOW)” (270957)
- Communication Theoretical Foundation of Wireless Nanonetworks (287112)
This publication has 14 references indexed in Scilit:
- Comprehensive Survey of Galvanic Coupling and Alternative Intra-Body Communication TechnologiesIEEE Communications Surveys & Tutorials, 2018
- In-Body and Off-Body Channel Modeling for Future Leadless Cardiac Pacemakers Based on Phantom and Animal ExperimentsIEEE Antennas and Wireless Propagation Letters, 2018
- Low-Power, High Data-Rate Digital Capsule Endoscopy Using Human Body CommunicationApplied Sciences, 2018
- Wireless Channel Modeling for Leadless Cardiac Pacemaker: Effects ofVentricular Blood Volume2018
- Galvanically Coupled Intrabody Communications for Medical Implants: A Unified Analytic ModelIEEE Transactions on Antennas and Propagation, 2016
- Multi-Path Model and Sensitivity Analysis for Galvanic Coupled Intra-Body Communication Through Layered TissueIEEE Transactions on Biomedical Circuits and Systems, 2015
- Investigation of Galvanic-Coupled Intrabody Communication Using the Human Body Circuit ModelIEEE Journal of Biomedical and Health Informatics, 2014
- Galvanic Coupling Enabling Wireless Implant CommunicationsIEEE Transactions on Instrumentation and Measurement, 2009
- High-resolution human anatomy models for advanced electromagnetic field computationsIEEE Transactions on Magnetics, 2002
- The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHzPhysics in Medicine & Biology, 1996