Incorporation of intraoperative computerized tomography in a newly developed spinal navigation technique
Authors:
Open Access
- 1 January 2000
- journal article
- research article
- Published by Taylor & Francis in Computer Aided Surgery
- Vol. 5 (1) , 18-27
Abstract
Objective: We report on the first successful incorporation of intraoperative computerized tomography in spinal navigation procedures. Materials and Methods: All operations were performed with the aid of a Tomoscan M mobile CT system (Philips Medical Systems, Eindhoven, The Netherlands). The system comprises a mobile gantry, a mobile patient examination table, and a mobile workstation. Three different navigation systems were used: the EasyGuide (Philips Medical Systems, Eindhoven, The Netherlands), the Surgical Tool Navigator (Zeiss, Oberkochen, Germany), and a prototype of an ultrasound navigation system developed by ourselves (TVS GbR, Chemnitz, Germany). All surgical operations were performed with the patient positioned on the mobile table of the CT system. Following dorsal preparation of the vertebral region, the surgeon implanted small titanium screws in the vertebrae to serve as fiducial markers. Image data acquisition and image-to-patient registration were performed after implantation of the marker screws. The pedicle screws were inserted using the navigation system, and the position of each implant was confirmed by intraoperative CT scans. Results: To date, 35 patients with various spine disorders have been operated upon using the technique described, and 161 pedicle screws have been inserted. There were no misplacements of pedicle screws. In three (1.9%) cases we observed lateral perforation (maximum 2 mm) of the lateral pedicle wall. We achieved a registration error of 0.85 mm (RMSE) ±0.42 mm (SD). Conclusions: The combination of intraoperative computerized tomography and spinal navigation allows easy navigation with a high application accuracy of 0.8 mm ± 0.4 mm (SD) at the target point (measured in experiments with a plastic spine model). The possibility of performing an intraoperative quality check on demand markedly improves the safety of these procedures.Keywords
This publication has 26 references indexed in Scilit:
- Experience with a New Multifunctional Articulated Instrument Holder in Minimally Invasive Navigated Neurosurgerymin - Minimally Invasive Neurosurgery, 1998
- The EasyGuide Neuro Image-Guided Surgery SystemNeurosurgery, 1997
- Frameless Stereotactic Guidance for Surgery of the Upper Cervical SpineNeurosurgery, 1997
- Verification of the position of pedicle screws in lumbar spinal fusionEuropean Spine Journal, 1997
- Infrared-Based Neuronavigation and Cortical Motor Stimulation in the Management of Central-Region TumorsStereotactic and Functional Neurosurgery, 1997
- Intraoperative Imaging of the BrainStereotactic and Functional Neurosurgery, 1996
- Computertomographische Lagekontrolle von Pedikelschrauben in der SkoliosechirurgieZeitschrift für Orthopädie und ihre Grenzgebiete, 1996
- Preliminary Clinical Experience with Linear Accelerator-based Spinal Stereotactic RadiosurgeryNeurosurgery, 1995
- Accuracy in Frame-Based and Frameless StereotaxyStereotactic and Functional Neurosurgery, 1993
- Atlantoaxial Mobility After Screw Fixation of the OdontoidJournal of Spinal Disorders, 1991