Ruby-based inorganic scintillation detectors for192Ir brachytherapy
- 14 October 2016
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 61 (21) , 7744-7764
Abstract
We tested the potential of ruby inorganic scintillation detectors (ISDs) for use in brachytherapy and investigated various unwanted luminescence properties that may compromise their accuracy. The ISDs were composed of a ruby crystal coupled to a poly(methyl methacrylate) fiber-optic cable and a charge-coupled device camera. The ISD also included a long-pass filter that was sandwiched between the ruby crystal and the fiber-optic cable. The long-pass filter prevented the Cerenkov and fluorescence background light (stem signal) induced in the fiber-optic cable from striking the ruby crystal, which generates unwanted photoluminescence rather than the desired radioluminescence. The relative contributions of the radioluminescence signal and the stem signal were quantified by exposing the ruby detectors to a high-dose-rate brachytherapy source. The photoluminescence signal was quantified by irradiating the fiber-optic cable with the detector volume shielded. Other experiments addressed time-dependent luminescence properties and compared the ISDs to commonly used organic scintillator detectors (BCF-12, BCF-60). When the brachytherapy source dwelled 0.5 cm away from the fiber-optic cable, the unwanted photoluminescence was reduced from >5% to <1% of the total signal as long as the ISD incorporated the long-pass filter. The stem signal was suppressed with a band-pass filter and was <3% as long as the source distance from the scintillator was 5% within 10 s from the onset of irradiation and after the source had retracted. The ruby-based ISDs generated signals of up to 20 times that of BCF-12-based detectors. The study presents solutions to unwanted luminescence properties of ruby-based ISDs for high-dose-rate brachytherapy. An optic filter should be sandwiched between the ruby crystal and the fiber-optic cable to suppress the photoluminescence. Furthermore, we recommend avoiding ruby crystals that exhibit significant time-dependent luminescence.Keywords
Funding Information
- National Cancer Institute (CA153824)
This publication has 38 references indexed in Scilit:
- On the nature of the light produced within PMMA optical light guides in scintillation fiber-optic dosimetryPhysics in Medicine & Biology, 2013
- Radioluminescence of red-emitting Eu-doped phosphors for fiberoptic dosimetryApplied Radiation and Isotopes, 2012
- In vivo dosimetry for gynaecological brachytherapy using a novel position sensitive radiation detector: Feasibility studyMedical Physics, 2012
- Fiber-coupled radioluminescence dosimetry with saturated Al 2 O 3 :C crystals: Characterization in 6 and 18 MV photon beamsRadiation Measurements, 2011
- Time‐resolved in vivo luminescence dosimetry for online error detection in pulsed dose‐rate brachytherapyMedical Physics, 2009
- Scintillating fiber dosimeter for radiation therapy acceleratorIEEE Transactions on Nuclear Science, 2002
- Cerenkov light generated in optical fibres and other light pipes irradiated by electron beamsPhysics in Medicine & Biology, 1992
- Optical properties of heavily doped rubyPhysica Status Solidi (a), 1972
- Energy Transfer in RubyPhysical Review B, 1967
- Cathodoluminescence of RubyJournal of Applied Physics, 1966