Abstract
Monte Carlo simulations play an essential role in many of the research activities on new scintillation detectors for X-ray-based medical imagers. The scintillation process, the transport of scintillation photons through the crystal(s) and X-photons interactions have a major influence on the detector performance; hence, considering all of these processes in the model is important to obtain accurate results. In this work, the behavior of crystalline zinc oxide as a scintillator detector under X-ray illumination is simulated by using the GEANT4 Monte Carlo simulation code. The optical and scintillation models of the GEANT4 are employed to investigate the optical response of zinc oxide as the X-ray detector in energies up to 100 keV (suitable for medical imaging applications). The results show that ZnO suffers from high self-absorption of the X-ray-generated optical photons and this problem strongly affects the performance of the ZnO as the X-ray scintillator detector especially for higher energies.

This publication has 13 references indexed in Scilit: