Dynamic trajectory‐based couch motion for improvement of radiation therapy trajectories in cranial SRT
- 1 May 2015
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 42 (5) , 2317-2325
Abstract
Purpose: To investigate potential improvement in external beam stereotactic radiation therapy plan quality for cranial cases using an optimized dynamic gantry and patient support couch motion trajectory, which could minimize exposure to sensitive healthy tissue.Methods: Anonymized patient anatomy and treatment plans of cranial cancer patients were used to quantify the geometric overlap between planning target volumes and organs‐at‐risk (OARs) based on their two‐dimensional projection from source to a plane at isocenter as a function of gantry and couch angle. Published dose constraints were then used as weighting factors for the OARs to generate a map of couch‐gantry coordinate space, indicating degree of overlap at each point in space. A couch‐gantry collision space was generated by direct measurement on a linear accelerator and couch using an anthropomorphic solid‐water phantom. A dynamic, fully customizable algorithm was written to generate a navigable ideal trajectory for the patient specific couch‐gantry space. The advanced algorithm can be used to balance the implementation of absolute minimum values of overlap with the clinical practicality of large‐scale couch motion and delivery time. Optimized cranial cancer treatment trajectories were compared to conventional treatment trajectories.Results: Comparison of optimized treatment trajectories with conventional treatment trajectories indicated an average decrease in mean dose to the OARs of 19% and an average decrease in maximum dose to the OARs of 12%. Degradation was seen for homogeneity index (6.14% ± 0.67%–5.48% ± 0.76%) and conformation number (0.82 ± 0.02–0.79 ± 0.02), but neither was statistically significant. Removal of OAR constraints from volumetric modulated arc therapy optimization reveals that reduction in dose to OARs is almost exclusively due to the optimized trajectory and not the OAR constraints.Conclusions: The authors’ study indicated that simultaneous couch and gantry motion during radiation therapy to minimize the geometrical overlap in the beams‐eye‐view of target volumes and the organs‐at‐risk can have an appreciable dose reduction to organs‐at‐risk.Keywords
This publication has 13 references indexed in Scilit:
- Dynamic electron arc radiotherapy (DEAR): a feasibility studyPhysics in Medicine & Biology, 2013
- Trajectory modulated prone breast irradiation: A LINAC-based technique combining intensity modulated delivery and motion of the couchRadiotherapy and Oncology, 2013
- Trajectory optimization for dynamic couch rotation during volumetric modulated arc radiotherapyPhysics in Medicine & Biology, 2013
- Simultaneous couch and gantry dynamic arc rotation (CG‐Darc) in the treatment of breast cancer with accelerated partial breast irradiation (APBI): a feasibility studyJournal of Applied Clinical Medical Physics, 2013
- Continuous Arc Rotation of the Couch Therapy for the Delivery of Accelerated Partial Breast Irradiation: A Treatment Planning AnalysisInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Choreographing Couch and Collimator in Volumetric Modulated Arc TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Radiation Dose–Volume Effects in the BrainInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2010
- Volumetric modulated arc therapy: IMRT in a single gantry arcMedical Physics, 2007
- A technique for the quantitative evaluation of dose distributionsMedical Physics, 1998
- A conformation number to quantify the degree of conformality in brachytherapy and external beam irradiation: Application to the prostateInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1997