Comparison of Unit Resolution Srm And Tof-Ms At 12,000 Mass Resolution For Quantitative Bioanalysis of 11 Steroids From Human Plasma
Authors:
- 12 March 2012
- journal article
- research article
- Published by Taylor & Francis in Bioanalysis
- Vol. 4 (5) , 555-563
Abstract
The use of high-resolution MS systems for quantitative bioanalysis is a growing field, even though a clear majority of bioanalytical methods are still based on MS/MS with triple quadrupole (QqQ) instrumentation. The recent advances in TOF-MS technology have provided increased linear range and a high selectivity of detection by increased mass resolution and mass accuracy, making these instruments attractive for quantitative analysis due to lack of a need for compound-specific detection reaction optimization and their capability to collect data for a high number of compounds by sensitive wide mass range data acquisition. Here, 11 steroids spiked to human plasma were analyzed by LC-MS using both a QqQ MS system and a TOF instrument operating at 12,000 mass resolution. Sample preparation was performed by hybrid SPE technology. The LOD were 0.5-5 and 0.5-20 ng/ml in plasma for all analytes with QqQ and TOF-MS detection, respectively. Although the results show wider linear range and slightly better sensitivity for most of the compounds with QqQ in comparison to TOF, acceptable performance was obtained for most of the compounds within the range of LOD to 2000 ng/ml (in plasma), this was also the case with LC-TOF-MS analysis. The main problem in TOF-MS analysis at 12,000 mass resolution from plasma was selectivity rather than sensitivity or linear range.Keywords
This publication has 20 references indexed in Scilit:
- A novel strategy for reducing phospholipids-based matrix effect in LC–ESI-MS bioanalysis by means of HybridSPEJournal of Pharmaceutical and Biomedical Analysis, 2009
- Ultra‐performance liquid chromatography/tandem mass spectrometry in high‐throughput detection, quantification and confirmation of anabolic steroids in equine plasmaRapid Communications in Mass Spectrometry, 2009
- Effect of eluent on the ionization process in liquid chromatography–mass spectrometryJournal of Chromatography A, 2009
- Quantitative measurements of corticosteroids in ex vivo samples using on-line SPE-LC/MS/MSJournal of Chromatography B, 2009
- Development of a liquid chromatography–tandem mass spectrometry method for the determination of 23 endogenous steroids in small quantities of primate urineJournal of Chromatography B, 2008
- A general screening method for doping agents in human urine by solid phase extraction and liquid chromatography/time-of-flight mass spectrometryAnalytica Chimica Acta, 2007
- Liquid-phase microextraction for sample preparation in analysis of unconjugated anabolic steroids in urineAnalytica Chimica Acta, 2006
- Screening of free 17-alkyl-substituted anabolic steroids in human urine by liquid chromatography–electrospray ionization tandem mass spectrometrySteroids, 2004
- Feasibility of a liquid‐phase microextraction sample clean‐up and liquid chromatographic/mass spectrometric screening method for selected anabolic steroid glucuronides in biological samplesJournal of Mass Spectrometry, 2003
- Electrospray collision-induced dissociation of testosterone and testosterone hydroxy analogsJournal of Mass Spectrometry, 1999