Repeatability of a fully automated swept-source optical coherence tomography biometer and agreement with a low coherence reflectometry biometer
Open Access
- 2 June 2023
- journal article
- research article
- Published by Springer Nature in Eye and Vision
- Vol. 10 (1) , 1-9
Abstract
Background: To evaluate the repeatability of a fully automated swept-source optical coherence tomography (SS-OCT) and its agreement with an optical low coherence reflectometry (OLCR) for several biometric parameters. Methods: In this study, 74 eyes of 74 patients were measured using the Eyestar 900 SS-OCT and Lenstar LS 900 OLCR. Flat keratometry (K1) and steep keratometry (K2), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), and axial length (AL) were measured three times with each device. The repeatability was analyzed with the intrasubject standard deviation, coefficient of variability (CoV), and coefficient of repeatability (CoR) for each instrument. The agreement between the instruments was evaluated with Bland-Altman analysis. Results: K1, K2 and CCT CoV values were < 0.2%, < 0.4% and < 0.55%, respectively. Higher CoV values were found for ACD and LT ranging from 0.56% to 1.74%. The lowest CoV values were found for the AL measurements (0.03% and 0.06% for the Eyestar 900 and the Lenstar LS 900, respectively). AL measurements provided the highest repeatability, measured with both CoV and CoR values, and the CCT was the parameter with the lowest repeatability. The CCT and LT measurements were statistically significant between the two biometers (P < 0.001). The interval of the limits of agreement was < 0.6 D for K1 and K2, 15.78 µm for CCT, 0.21 mm for ACD, 0.34 mm for LT, and 0.08 mm for AL. Conclusions: Both biometers provide repeatable measurements for the different parameters analyzed and can be used interchangeably.Keywords
Funding Information
- Karolinska Institute
This publication has 20 references indexed in Scilit:
- Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography–based biometer versus partial coherence interferometry and optical low-coherence reflectometryJournal of Cataract & Refractive Surgery, 2016
- Precision (repeatability and reproducibility) studies and sample-size calculationJournal of Cataract & Refractive Surgery, 2015
- Reproducibility of a Long-Range Swept-Source Optical Coherence Tomography Ocular Biometry System and Comparison with Clinical BiometersOphthalmology, 2013
- Biometric measurements in highly myopic eyesJournal of Cataract & Refractive Surgery, 2013
- Evaluation of the repeatability of the Lenstar and comparison with two other non‐contact biometric devices in myopesClinical and Experimental Optometry, 2013
- Scheimpflug–Placido topographer and optical low-coherence reflectometry biometer: Repeatability and agreementJournal of Cataract & Refractive Surgery, 2012
- Clinically relevant biometryCurrent Opinion in Ophthalmology, 2012
- Repeatability and Reproducibility of Biometry and Keratometry Measurements Using a Noncontact Optical Low-Coherence Reflectometer and KeratometerAmerican Journal of Ophthalmology, 2012
- Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmologyOphthalmic and Physiological Optics, 2011
- Comparison of Predictions Made by the Intraocular Lens Master and Ultrasound BiometryArchives of Ophthalmology (1950), 2008