Multichannel bolometer for radiation measurements on the TCA tokamak
Authors:
- 1 October 1986
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 57 (10) , 2449-2454
Abstract
A multichannel radiation bolometer has been developed for the Tokamak Chauffage Alfvén (TCA) tokamak. It has 16 equally spaced chords that view the plasma through a narrow horizontal slit. Almost an entire vertical plasma cross section can be observed. The bolometer operates on the basis of a semiconducting element which serves as a temperature-dependent resistance. A new electronic circuit has been developed which takes advantage of the semiconductor characteristics of the detector by using feedback techniques. Measurements made with this instrument are discussed.Keywords
This publication has 11 references indexed in Scilit:
- Bolometric diagnostics in JETReview of Scientific Instruments, 1985
- Performance of the Tokamak Fusion Test Reactor bolometersReview of Scientific Instruments, 1985
- A new metal resistor bolometer for measuring vacuum ultraviolet and soft x radiationJournal of Applied Physics, 1984
- Measurements of electron and ion heating by Alfven waves in the TCA TokamakPlasma Physics, 1983
- Radiation losses and global energy balance for Ohmically heated discharges in ASDEXNuclear Fusion, 1982
- Bolometer for measurements on high-temperature plasmasReview of Scientific Instruments, 1982
- Ionization equilibrium and radiative cooling of a high temperature plasmaJournal of Quantitative Spectroscopy and Radiative Transfer, 1978
- Measurement of the energy balance in ATC tokamakNuclear Fusion, 1975
- A thin film thermopile for neutral particle beam measurementsReview of Scientific Instruments, 1974
- Reflectivities of Evaporated Metal Films in the Near and Far UltravioletPhysical Review B, 1939