Repeatability of Corneal Epithelial Thickness Measurements Using Fourier-Domain Optical Coherence Tomography in Normal and Post-LASIK Eyes
Authors:
- 1 December 2013
- journal article
- research article
- Published by Wolters Kluwer Health in Cornea
- Vol. 32 (12) , 1544-1548
Abstract
Purpose: To evaluate the repeatability of corneal epithelial thickness (ET) and corneal thickness (CT) measurements in normal eyes and laser in situ keratomileusis (LASIK)-treated eyes using optical coherence tomography (RTVue system). Methods: In 35 right eyes of 35 normal subjects and 45 right eyes of 45 subjects who underwent a myopic LASIK, corneal ET and CT were evaluated in 17 areas: (1) 1 central zone within a 0- to 2.0-mm diameter, (2) 8 paracentral zones from a 2.0- to 5.0-mm diameter, and (3) 8 peripheral zones from a 5.0- to 6.0-mm diameter. The repeatability was assessed using within-subject standard deviation (SD), coefficient of variation, and intraclass correlation coefficient. Results: At the central and paracentral zones, respectively, the SD values were 0.7 μm and 0.6 to 0.9 μm in normal eyes and 0.7 μm and 0.8 to 1.7 μm in LASIK-treated eyes for ET, and 1.0 μm and 2.8 to 4.6 μm in normal eyes and 1.3 μm and 4.0 to 4.8 μm in LASIK-treated eyes for CT. At the peripheral zones, in normal and LASIK-treated eyes, respectively, the SD values ranged from 0.8 to 1.2 μm and 1.4 to 2.2 μm for ET, and 4.1 to 6.4 μm and 6.0 to 9.1 μm for CT. The coefficient of variation values were low and intraclass correlation coefficient values were high in both groups for both ET and CT measurements. Conclusions: The Optical coherence tomography produced excellent repeatability, especially at the central and paracentral zones up to a 5-mm diameter for both corneal ET and CT measurements.Keywords
This publication has 21 references indexed in Scilit:
- Corneal Epithelial Thickness Mapping by Fourier-Domain Optical Coherence Tomography in Normal and Keratoconic EyesOphthalmology, 2012
- Corneal thickness mapping by 3D swept‐source anterior segment optical coherence tomographyActa Ophthalmologica, 2012
- Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness MeasurementYonsei Medical Journal, 2012
- Repeatability and reproducibility of anterior ocular biometric measurements with 2-dimensional and 3-dimensional optical coherence tomographyJournal of Cataract & Refractive Surgery, 2010
- Pachymetric mapping with Fourier-domain optical coherence tomographyJournal of Cataract & Refractive Surgery, 2010
- Repeatability and Reproducibility of Pachymetric Mapping with Visante Anterior Segment–Optical Coherence TomographyInvestigative Ophthalmology & Visual Science, 2007
- The Repeatability of Corneal and Corneal Epithelial Thickness Measurements Using Optical Coherence TomographyOptometry and Vision Science, 2006
- Effect of myopic laser in situ keratomileusis on epithelial and stromal thicknessOphthalmology, 2002
- Epithelial and corneal thickness measurements by in vivo confocal microscopy through focusing (CMTF)Current Eye Research, 1997
- Statistics Notes: Measurement errorBMJ, 1996