Overcoming Ru-99 Isobaric Overlap via C2H4 as a Reaction Gas for Tc-99 Accurate Measurement in Large-Volume Seawater Samples by Triple Quadrupole ICP−MS with Mass-Shift Mode
Authors:
- 17 November 2026
- journal article
- Published by American Chemical Society (ACS) in Analytical Chemistry
Abstract
Accurate determination of ultra-trace level 99Tc in environmental samples is primarily hindered by the interferences of 98Mo and 99Ru. For the “elimination” of the 99Ru isobaric overlap during the mass spectrometry measurement, the commonly-used method of mathematical correction based on a known isotopic ratio of Ru (e.g., 99Ru/102Ru) could lead to a high measurement uncertainty of 99Tc. In this work, a direct and accurate analytical method of 99Tc in large-volume seawater samples (>20 L) was developed using two TEVA resin columns in combination with ICP−MS/MS in He−C2H4 mode. Both Fe(OH)2 and Ti(OH)3 were used as co-precipitants for the pre-concentration of 99Tc in large-volume seawater samples, and then, 99Tc was purified by two serial TEVA resin columns with the high decontamination factors of Mo (5×105) and Ru (1×105). Based on the high reactivity of C2H4, the residual interferences of 98Mo and 99Ru were completely removed in the mass-shift mode for the first time. Excellent decontamination factors of 98Mo (1×1012) and 99Ru (1.6×109) were achieved in the whole procedure, which was 3−4 orders of magnitude higher than these of traditional methods. The detection limit of 0.015 pg/L for 99Tc was obtained, and the whole analysis time was less than 20 h per sample. This method is expected to directly and accurately measure ultra-trace level 99Tc in large-volume seawater samples without any mathematical correction of 99Ru.Funding Information
- Science and Technology Program of Gansu Province (26YFWA004)
- National Natural Science Foundation of China (12175201)
- National Natural Science Foundation of China (W2412046)
This publication has 37 references indexed in Scilit:
- Quantification of [99Tc]TcO4- in urine by means of anion-exchange chromatography–aerosol desolvation nebulization–inductively coupled plasma–mass spectrometryAnalytical and Bioanalytical Chemistry, 2024
- Selectivity analysis of 99Tc determination by LSC in the field of nuclear decommissioningJournal of Radioanalytical and Nuclear Chemistry, 2022
- Simultaneous separation of actinides and technetium from large volumes of natural water for their determinationJournal of Radioanalytical and Nuclear Chemistry, 2022
- Liquid scintillation counting for determination of radionuclides in environmental and nuclear applicationJournal of Radioanalytical and Nuclear Chemistry, 2018
- Preconcentration and separation of 99Tc in groundwater by using TEVA resinJournal of Radioanalytical and Nuclear Chemistry, 2017
- Determination of technetium-99 in environmental samples: A reviewAnalytica Chimica Acta, 2012
- Technetium2000
- Separation and preconcentration of actinides by extraction chromatography using a supported liquid anion exchanger: application to the characterization of high-level nuclear waste solutionsAnalytica Chimica Acta, 1995
- Coprecipitation with tin(IV) hydroxide followed by removal of tin carrier for the determination of trace heavy metals by graphite-furnace atomic absorption spectrometryAnalytica Chimica Acta, 1995
- Determination of technetium-99 from the aspect of environmental radioactivityJournal of Radioanalytical and Nuclear Chemistry, 1990