Comparison of Two Multifocal Intraocular Lens Designs That Differ Only in Near Add
Authors:
- Emilio PedrottiEPEmilio Pedrotti
- Rodolfo MastropasquaRodolfo Mastropasqua
- Mattia PassilongoMPMattia Passilongo
- Graziella ParisiGPGraziella Parisi
- Ivan MarchesoniIMIvan Marchesoni
- Giorgio MarchiniGMGiorgio Marchini
- 1 November 2014
- journal article
- research article
- Published by SLACK, Inc. in Journal of Refractive Surgery
- Vol. 30 (11) , 754-760
Abstract
PURPOSE: To evaluate monocular functional outcomes after the implantation of the AcrySof ReSTOR SN6AD2 intraocular lens (IOL) (+2.5 diopters [D] near add) (Alcon Laboratories, Inc., Fort Worth, TX) and the AcrySof ReSTOR SN6AD1 IOL (+3.0 D near add) (Alcon Laboratories, Inc.). METHODS: This prospective, comparative, nonrandomized single-blind observational study comprised 62 eyes of 62 patients that underwent phacoemulsification and implantation of a multifocal IOL: SN6AD2 (31 eyes) (+2.5 group) and SN6AD1 (31 eyes) (+3.0 group). Twelve months after surgery, monocular near (30 and 40 cm), intermediate (50, 60, and 70 cm), and distance (4 m) visual acuity were evaluated with the internal root mean square and modulation transfer function. Both parameters were evaluated at the 4- and 6-mm pupil sizes. RESULTS: No statistical differences at 4 m were found between the groups. The +2.5 group obtained better performances at all intermediate distances (50 cm, 0.23 ± 0.14 vs 0.32 ± 0.13; 60 and 70 cm, 0.21 ± 0.10 vs 0.41 ± 0.14 and 0.24 ± 0.10 vs 0.53 ± 0.17, respectively), whereas the near visual acuity was better for the +3.0 group (30 and 40 cm, 0.36 ± 0.18 vs 0.14 ± 0.09 and 0.36 ± 0.19 vs 0.17 ± 0.07, respectively). The root mean square was lower for the +2.5 group compared to the +3.0 group, whereas the modulation transfer function showed overlapping results between the two models. CONCLUSIONS: Both IOL models showed good results in distance vision; the +2.5 D IOL seemed to provide better intermediate vision than the +3.0 D IOL. For near vision, the +3.0 D model performed better than the +2.5 D model. [ J Refract Surg . 2014;30(11):754–760.]Keywords
This publication has 26 references indexed in Scilit:
- Multifocal intraocular lens with a surface-embedded near section: Short-term clinical outcomesJournal of Cataract & Refractive Surgery, 2011
- Visual outcomes and optical performance of a monofocal intraocular lens and a new-generation multifocal intraocular lensJournal of Cataract & Refractive Surgery, 2011
- Visual outcome and patient satisfaction after multifocal intraocular lens implantation: Aspheric versus spherical designJournal of Cataract & Refractive Surgery, 2010
- Visual outcomes and optical performance with a monofocal intraocular lens and a new-generation single-optic accommodating intraocular lensJournal of Cataract & Refractive Surgery, 2010
- Optical quality of foldable monofocal intraocular lenses before and after injectionJournal of Cataract & Refractive Surgery, 2009
- Visual outcome comparison of unilateral versus bilateral implantation of apodized diffractive multifocal intraocular lenses after cataract extractionJournal of Cataract & Refractive Surgery, 2009
- Apodized diffractive versus refractive multifocal intraocular lenses: Optical and visual evaluationJournal of Cataract & Refractive Surgery, 2008
- History and development of the apodized diffractive intraocular lensJournal of Cataract & Refractive Surgery, 2006
- Multifocal intraocular lensesCurrent Opinion in Ophthalmology, 2005
- Cataract extraction with multifocal intraocular lens implantationOphthalmology, 2000