Resolution recovery on list mode MLEM reconstruction for Dynamic Cardiac SPECT system
Authors:
Preprint
Abstract
The Dynamic Cardiac SPECT (DC-SPECT) system is being developed at the Massachusetts General Hospital, featuring a static cardio focus asymmetrical geometry enabling simultaneous high resolution and high sensitivity imaging. Among 14 design iterations of the DC-SPECT with varying number of detector heads, system sensitivity and system resolution, the current version being fabricated features 10 mm FWHM geometrical resolution (without resolution recovery) and 0.07% sensitivity; this is 1.5x resolution gain and 7x sensitivity gain compared to a conventional dual head gamma camera. This work presents improvement in imaging resolution by implementing a spatially variant point spread function (SV-PSF) with list mode MLEM reconstruction. A resolution recovery method by PSF deconvolution is validated on list mode MLEM reconstruction for DC-SPECT. A spatial invariant PSF is included as an additional test to show the influence of the accuracy of PSF modelling on the reconstructed image quality. We compare the MLEM reconstruction with and without PSF deconvolution; an analytic model is used for the calculation of system response, and the results are compared to Monte Carlo (MC) based methods. Results show that with PSF modelling applied, the quality of the reconstructed image is improved, and the DC-SPECT system can achieve a 4.5 mm central spatial resolution with average 795 counts/(s*Mbq). The results show substantial improvement over the gold standard GE Discovery 570c performance (spatial resolution 7 mm with an average 460 counts/s*MBq, central resolution 5.8 mm). The impact of PSF deconvolution is significant, and the improvements of the reconstructed image quality is more evident compared to MC simulated system matrix with the same sampling size as in simulation.Keywords
All Related Versions
This publication has 12 references indexed in Scilit:
- Impact of iterative reconstruction with resolution recovery in myocardial perfusion SPECT: phantom and clinical studiesScientific Reports, 2019
- Impact of Resolution Recovery in Quantitative 99mTc SPECT/CT Cardiac Phantom StudiesJournal of Medical Imaging and Radiation Sciences, 2019
- Light Spread Manipulation in Scintillators Using Laser Induced Optical BarriersIEEE Transactions on Nuclear Science, 2018
- GATE simulation of a high-performance stationary SPECT system for cardiac imaging2016
- Improved image quality in pinhole SPECT by accurate modeling of the point spread function in low magnification systemsMedical Physics, 2015
- Comparative analysis of iterative reconstruction algorithms with resolution recovery for cardiac SPECT studies. A multi-center phantom studyJournal of Nuclear Cardiology, 2014
- A System Calibration and Fast Iterative Reconstruction Method for Next-Generation SPECT ImagersIEEE Transactions on Nuclear Science, 2012
- EM algorithm system modeling by image-space techniques for PET reconstructionIEEE Transactions on Nuclear Science, 2003
- Improved SPECT quantitation using fully three-dimensional iterative spatially variant scatter response compensationIEEE Transactions on Medical Imaging, 1996
- Noise and Edge Artifacts in Maximum-Likelihood Reconstructions for Emission TomographyIEEE Transactions on Medical Imaging, 1987