Assessment of uncertainties for measurements of total near-normal emissivity of low-emissivity foils with an industrial emissometer
Open Access
- 15 June 2021
- journal article
- research article
- Published by Copernicus GmbH in Journal of Sensors and Sensor Systems
- Vol. 10 (1) , 135-152
Abstract
The TIR100-2 emissometer (manufactured by Inglas GmbH & Co.KG) is an emissivity measurement device used by several producers of thermal insulation products for buildings and by some organizations certifying performance of insulation products. A comparison of emissivity measurements on low-emissivity foils involving different measurement techniques, including the TIR100-2 emissometer, gave widely dispersed results; the discrepancies were not explained. The metrological performance of the TIR100-2 emissometer and the uncertainties for measurement on reflective foils was not known, which could be detrimental to users. In order to quantify the performance of TIR100-2 devices for measurement of total near-normal emissivity of low-emissivity foils, the Laboratoire National de Métrologie et d'Essais (LNE) analyzed in detail the measuring principle and listed the associated assumptions and uncertainty sources. A TIR100-2 emissometer actually measures the reflectance and, for opaque materials, the emissivity is calculated from the measured reflectance. The parameters analyzed experimentally are the temperature stability and uniformity of the thermal radiation source, the emissivity of the radiation source, the response function linearity and the spectral sensitivity of the radiometric detection system measuring the reflected radiation, the size of the measurement area, and the measurement repeatability and reproducibility. A detailed uncertainty budget was established. The uncertainty sources taken into account are the uncertainties of the emissivities of the two calibrated standards used for calibration, the stability and uniformity of the radiation source temperature, the non-linearity and the spectral sensitivity of the radiometric detection system, the specific measurement condition related to the radiation source temperature, the uncertainties related to the temperatures of the standards and the sample, the noises on results, and the non-homogeneity in emissivity of the tested material. The combined measurement uncertainty was calculated for different types of reflective foils; the expanded uncertainty is around 0.03 for total near-normal emissivity measurements on smooth low-emissivity foils. A measurement campaign on five types of low-emissivity foils, involving four TIR100-2 emissometers, and a comparison to a primary reference setup at the Physikalisch-Technische Bundesanstalt (PTB) confirmed the uncertainties assessed.Keywords
Funding Information
- European Metrology Programme for Innovation and Research (JRP 16NRM06 EMIRIM)
This publication has 7 references indexed in Scilit:
- Characterization, calibration and validation of an industrial emissometerJournal of Sensors and Sensor Systems, 2019
- Methods for quantitative infrared directional-hemispherical and diffuse reflectance measurements using an FTIR and a commercial integrating sphereApplied Optics, 2018
- Développement d’un réflectomètre absolu pour l’étalonnage de miroirs dans l’infrarougeRevue française de métrologie, 2014
- The determination of the uncertainties of spectral emissivity measurements in air at the PTBMetrologia, 2010
- Integrating-sphere system and method for absolute measurement of transmittance, reflectance, and absorptance of specular samplesApplied Optics, 2001
- A correlation between normal and hemispherical emissivity of low-emissivity coatings on glassInternational Communications in Heat and Mass Transfer, 1987
- Correction methods for integrating‐sphere measurement of hemispherical reflectanceColor Research & Application, 1986