Radiation dose vs. image quality for low-dose CT protocols of the head for maxillofacial surgery and oral implant planning
Authors:
- M. LoubeleMLM. Loubele
- R. JacobsRJR. Jacobs
- F. MaesFMF. Maes
- F. SchutyserFSF. Schutyser
- D. DebaveyeDDD. Debaveye
- R. BogaertsRBR. Bogaerts
- W. CoudyzerWCW. Coudyzer
- D. VandermeulenDVD. Vandermeulen
- J. van CleynenbreugelJCJ. van Cleynenbreugel
- G. MarchalGMG. Marchal
- P. SuetensPSP. Suetens
- 1 December 2005
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 117 (1-3) , 211-216
Abstract
The goal of this study was to determine the acquisition parameters for a low-dose multi-slice CT protocol and to compare the effective dose and the image quality of this low-dose protocol with the image quality of a clinical multi-slice CT protocol, routinely used for visualisation of the head. The low-dose protocol was derived from a clinical multi-slice CT protocol by lowering mA s and kV and increasing the pitch. The low-dose protocol yielded a dose reduction from 1.5 to 0.18 mSv for a multi-slice CT scan of the whole head, whereas noise in the low-dose CT images was increased. For bone segmentation, noise could be reduced by use of a non-linear edge preserving smoothing filter. Tests on ESP and skull phantom indicated that the accuracy of the measurements on low-dose CT is acceptable for image-based planning of maxillofacial and oral implant surgery, reducing the dose by a factor of 8.Keywords
This publication has 13 references indexed in Scilit:
- Spiral CT and radiation doseEuropean Journal of Radiology, 2003
- Head and neck imaging with MDCTEuropean Journal of Radiology, 2003
- Radiation doses of collimatedvsnon-collimated cephalometric exposuresDentomaxillofacial Radiology, 2003
- Low-dose CT of the paranasal sinuses with eye lens protection: effect on image quality and radiation doseEuropean Radiology, 2002
- Registration of 3D Photographs with Spiral CT Images for Soft Tissue Simulation in Maxillofacial Surgery2001
- Robust 3D Segmentation of Anatomical Structures with Level SetsLecture Notes in Computer Science, 2000
- Accuracy limits for the determination of cortical width and density: the influence of object size and CT imaging parametersPhysics in Medicine & Biology, 1999
- Multimodality image registration by maximization of mutual informationIEEE Transactions on Medical Imaging, 1997
- The European Spine Phantom — a tool for standardization and quality control in spinal bone mineral measurements by DXA and QCTEuropean Journal of Radiology, 1995
- Marching cubes: A high resolution 3D surface construction algorithmACM SIGGRAPH Computer Graphics, 1987