In Vitro Measurement of Contact Pressure Applied to a Model Vessel Wall during Balloon Dilation by Using a Film-Type Sensor
Open Access
- 1 January 2022
- journal article
- research article
- Published by The Japanese Society for Neuroendovascular Therapy in Journal of Neuroendovascular Therapy
- Vol. 16 (4) , 192-197
Abstract
Objective: As an important evaluation index of vascular damage, the study aims to clarify the value of contact pressure applied to blood vessels and how it changes with respect to balloon pressure during balloon dilation. Methods: The contact pressure was evaluated through an in vitro measurement system using a model tube with almost the same elastic modulus as the blood vessel wall and our film-type pressure sensor. A poly (vinyl alcohol) hydrogel tube with almost the same elastic modulus was fabricated as the model vessel. The film-type sensor was inserted between the balloon catheter and the model vessel, and the balloon was dilated. Results: The contact pressure applied to the blood vessel was less than 10% of the balloon pressure, and the increase in contact pressure was less than 1% of the increase in balloon pressure (8 to 14 atm). Moreover, the contact pressure and its increase were larger in the model with a high elastic modulus. Conclusion: The contact pressure to expand the soft vessel model was not high, and the balloon pressure almost appeared to act on the expansion of the balloon itself. Our experiment using variable stiffness vessel models containing film-type sensors showed that the contact pressure acting on the vessel wall tended to increase as the wall became harder, even when the nominal diameter of the balloon was almost identical to the vessel. Our results can be clinically interpreted: when a vessel is stiff, the high-pressure inflation may rupture it even if its nominal diameter is identical to the diameter of the vessel.Keywords
This publication has 13 references indexed in Scilit:
- Development of thin and flexible contact pressure sensing system for high spatial resolution measurementsSensors and Actuators A: Physical, 2017
- Fiberoptic Contact‐Force Sensing Electrophysiological Catheters: How Precise Is the Technology?Journal of Cardiovascular Electrophysiology, 2016
- A computational study of stent performance by considering vessel anisotropy and residual stressesMaterials Science and Engineering: C, 2016
- Factors affecting dilation force in balloon dilation of severe esophageal strictures: an experiment using an artificial stricture modelSurgical Endoscopy, 2016
- Discordance of the Areas of Peak Wall Shear Stress and Tissue Stress in Coronary Artery Plaques as Revealed by Fluid-Structure Interaction Finite Element AnalysisInternational Heart Journal, 2013
- Observation of local elastic distribution in aortic tissues under static strain condition by use of a scanning haptic microscopeJournal of Artificial Organs, 2012
- MEMS tri-axial force sensor with an integrated mechanical stopper for guidewire applicationsMicrosystem Technologies, 2012
- Soft tissue elastometerMedical Engineering & Physics, 2007
- Mechanisms of acute lumen gain following cutting balloon angioplasty in calcified and noncalcified lesions: An intravascular ultrasound studyCatheterization and Cardiovascular Interventions, 2002
- Stiffness and elastic behavior of human intracranial and extracranial arteriesJournal of Biomechanics, 1980