Biomechanical comparison of modified Calcanail system with plating fixation in intra-articular calcaneal fracture: A finite element analysis
- 20 June 2019
- journal article
- research article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 70 (1) , 55-61
Abstract
The calcanail system is a novel intramedullary approach for calcaneal fractures but is believed to be insufficient to treat complex fractures. We propose a modified Calcanail technique by adding a transfixation screw to improve stability. The aim of this study was to evaluate the biomechanical stability of the modified Calcanail system and compare it with the traditional Calcanail system and plate fixation. A Sanders type-IIIAB calcaneal fracture model was built and simulated fixation with the three implants. A vertical loading of 700 N was applied to the subtalar joint surfaces, and the posteroinferior calcaneal tuberosity was fixed. Construct stiffness, fracture migration, and von Mises stress were assessed. The results showed the modified Calcanail system demonstrated the highest construct stiffness, smallest migration, and lowest von Mises stress among three fixations. The study suggested that the modified Calcanail system can provide comparatively sufficient stability, that makes it preferable to treat complex calcaneal fractures.Keywords
Funding Information
- Hong Kong Polytechnic University (PolyU152216/14E)
- Hong Kong Polytechnic University (PolyU15200215)
- NSFC (11732015)
- Shanghai Municipal Health Burean (201840137)
This publication has 19 references indexed in Scilit:
- Interlocking Nailing Versus Interlocking Plating in Intra-articular Calcaneal FracturesFoot & Ankle International, 2016
- Finite element analysis of three commonly used external fixation devices for treating Type III pilon fracturesMedical Engineering & Physics, 2014
- Finite element comparison of retrograde intramedullary nailing and locking plate fixation with/without an intramedullary allograft for distal femur fracture following total knee arthroplastyThe Knee, 2014
- Primary stability of an intramedullary calcaneal nail and an angular stable calcaneal plate in a biomechanical testing model of intraarticular calcaneal fractureInjury, 2014
- Biomechanical Stability of Intramedullary Technique for Fixation of Joint Depressed Calcaneus FractureFoot & Ankle International, 2010
- Experimental identification and analytical modelling of human walking forces: Literature reviewJournal of Sound and Vibration, 2009
- Percutaneous Screw Configuration versus Perimeter Plating of Calcaneus Fractures: A Cadaver StudyFoot & Ankle International, 2008
- The characteristics of a hydroxyapatite–chitosan–PMMA bone cementBiomaterials, 2004
- Calcaneus fractures: facts, controversies and recent developmentsInjury, 2004
- Effect of dynamization on gap healing of diaphyseal fractures under external fixationClinical Biomechanics, 1995