Feasibility of absolute cerebral tissue oxygen saturation during cardiopulmonary resuscitation
Open Access
- 1 March 2013
- journal article
- research article
- Published by Springer Nature in Critical Care
- Vol. 17 (2) , R36
Abstract
Introduction: Current monitoring during cardiopulmonary resuscitation (CPR) is limited to clinical observation of consciousness, breathing pattern and presence of a pulse. At the same time, the adequacy of cerebral oxygenation during CPR is critical for neurological outcome and thus survival. Cerebral oximetry, based on near-infrared spectroscopy (NIRS), provides a measure of brain oxygen saturation. Therefore, we examined the feasibility of using NIRS during CPR. Methods: Recent technologies (FORE-SIGHT™ and EQUANOX™) enable the monitoring of absolute cerebral tissue oxygen saturation (SctO2) values without the need for pre-calibration. We tested both FORE-SIGHT™ (five patients) and EQUANOX Advance™ (nine patients) technologies in the in-hospital as well as the out-of-hospital CPR setting. In this observational study, values were not utilized in any treatment protocol or therapeutic decision. An independent t-test was used for statistical analysis. Results: Our data demonstrate the feasibility of both technologies to measure cerebral oxygen saturation during CPR. With the continuous, pulseless near-infrared wave analysis of both FORE-SIGHT™ and EQUANOX™ technology, we obtained SctO2 values in the absence of spontaneous circulation. Both technologies were able to assess the efficacy of CPR efforts: improved resuscitation efforts (improved quality of chest compressions with switch of caregivers) resulted in higher SctO2 values. Until now, the ability of CPR to provide adequate tissue oxygenation was difficult to quantify or to assess clinically due to a lack of specific technology. With both technologies, any change in hemodynamics (for example, ventricular fibrillation) results in a reciprocal change in SctO2. In some patients, a sudden drop in SctO2 was the first warning sign of reoccurring ventricular fibrillation. Conclusions: Both the FORE-SIGHT™ and EQUANOX™ technology allow non-invasive monitoring of the cerebral oxygen saturation during CPR. Moreover, changes in SctO2 values might be used to monitor the efficacy of CPR efforts.Keywords
This publication has 20 references indexed in Scilit:
- European Resuscitation Council Guidelines for Resuscitation 2010 Section 1. Executive summaryResuscitation, 2010
- The association of non-invasive cerebral and mixed venous oxygen saturation during cardiopulmonary resuscitationInteractive CardioVascular and Thoracic Surgery, 2010
- Recent Advances in Application of Cerebral Oximetry in Adult Cardiovascular SurgerySeminars in Cardiothoracic and Vascular Anesthesia, 2008
- Validation of the Cas Neonatal NIRS System by Monitoring VV-ECMO Patients2005
- Cerebral oximetry in out-of-hospital cardiac arrest: standard CPR rarely provides detectable hemoglobin–oxygen saturation to the frontal cortexResuscitation, 2004
- Principles, Techniques, and Limitations of Near Infrared SpectroscopyCanadian Journal of Applied Physiology, 2004
- Arterial and Venous Contributions to Near-infrared Cerebral OximetryAnesthesiology, 2000
- Near-infrared oximetry of the brainProgress in Neurobiology, 1999
- Near infrared spectroscopyBritish Journal of Anaesthesia, 1999
- Noninvasive, Infrared Monitoring of Cerebral and Myocardial Oxygen Sufficiency and Circulatory ParametersScience, 1977