Characterization of the Targeting Accuracy of a Neuronavigation-Guided Transcranial FUS System In Vitro, In Vivo, and In Silico
- 14 November 2022
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 70 (5) , 1528-1538
Abstract
Focused ultrasound (FUS)-enabled liquid biopsy (sonobiopsy) is an emerging technique for the noninvasive and spatiotemporally controlled diagnosis of brain cancer by inducing blood-brain barrier (BBB) disruption to release brain tumor-specific biomarkers into the blood circulation. The feasibility, safety, and efficacy of sonobiopsy were demonstrated in both small and large animal models using magnetic resonance-guided FUS devices. However, the high cost and complex operation of magnetic resonance-guided FUS devices limit the future broad application of sonobiopsy in the clinic. In this study, a neuronavigation-guided sonobiopsy device is developed and its targeting accuracy is characterized in vitro , in vivo , and in silico . The sonobiopsy device integrated a commercially available neuronavigation system (BrainSight) with a nimble, lightweight FUS transducer. Its targeting accuracy was characterized in vitro in a water tank using a hydrophone. The performance of the device in BBB disruption was verified in vivo using a pig model, and the targeting accuracy was quantified by measuring the offset between the target and the actual locations of BBB opening. The feasibility of the FUS device in targeting glioblastoma (GBM) tumors was evaluated in silico using numerical simulation by the k-Wave toolbox in glioblastoma patients. It was found that the targeting accuracy of the neuronavigation-guided sonobiopsy device was 1.7 ± 0.8 mm as measured in the water tank. The neuronavigation-guided FUS device successfully induced BBB disruption in pigs with a targeting accuracy of 3.3 ± 1.4 mm. The targeting accuracy of the FUS transducer at the GBM tumor was 5.5 ± 4.9 mm. Age, sex, and incident locations were found to be not correlated with the targeting accuracy in GBM patients. This study demonstrated that the developed neuronavigation-guided FUS device could target the brain with a high spatial targeting accuracy, paving the foundation for its application in the clinic.Keywords
Funding Information
- National Institutes of Health (R01EB027223, R01EB030102, R01MH116981, UG3MH126861)
- Office of Naval Research (N00014-19-1-2335)
This publication has 39 references indexed in Scilit:
- The “Liquid Biopsy”: the Role of Circulating DNA and RNA in Central Nervous System TumorsCurrent Neurology and Neuroscience Reports, 2016
- Detection of Circulating Tumor DNA in Early- and Late-Stage Human MalignanciesScience Translational Medicine, 2014
- Targeting accuracy of transcranial magnetic resonance–guided high-intensity focused ultrasound brain therapy: a fresh cadaver modelJournal of Neurosurgery, 2013
- Correction: MR-guided focused ultrasound technique in functional neurosurgery: targeting accuracyJournal of Therapeutic Ultrasound, 2013
- Neuronavigation-Guided Focused Ultrasound-Induced Blood-Brain Barrier Opening: A Preliminary Study in SwineAmerican Journal of Neuroradiology, 2012
- k-Wave: MATLAB toolbox for the simulation and reconstruction of photoacoustic wave fieldsJournal of Biomedical Optics, 2010
- Simulation of Intracranial Acoustic Fields in Clinical Trials of SonothrombolysisUltrasound in Medicine & Biology, 2009
- A non-invasive method for focusing ultrasound through the human skullPhysics in Medicine & Biology, 2002
- The calibration of CT Hounsfield units for radiotherapy treatment planningPhysics in Medicine & Biology, 1996
- Speed of Sound in Pure WaterThe Journal of the Acoustical Society of America, 1972