Collaborative guarded-hot-plate tests between the Laboratoire national de métrologie et d’essais and the National Institute of Standards and Technology
- 20 January 2017
- journal article
- research article
- Published by IOP Publishing in Metrologia
- Vol. 54 (1) , 113-128
Abstract
A collaborative study to compare the long-term measurement performance between guarded-hot-plate facilities at the Laboratoire national de métrologie et d’essais (LNE) in France and the National Institute of Standards and Technology (NIST) in the United States is presented. Thermal conductivity data were compiled from three international comparisons organized from 1997 to 2014. Measurements were conducted in accordance with standardized test methods (ISO 8302 or ASTM C 177) over a temperature range from 280 K to 320 K. Nine thermal insulating materials (either mineral fiber or expanded polystyrene) were examined covering broad ranges of bulk densities (13 kg · m −3 –200 kg · m −3 ) and thicknesses (13 mm–70 mm). A different set of specimens was utilized for each comparison. Results of this study indicate that, over a 17 year interval, the majority of test data from LNE and NIST agreed to within ±1.0%, or less, for mineral fiber materials and to within ±0.5%, or less, for expanded polystyrene. The long-term variability limit of 1% between the two laboratories is in good agreement with their current measurement uncertainties. Regression coefficients and their standard uncertainties for a straight-line model relating thermal conductivity to temperature from 280 K to 320 K were computed by material and laboratory. Graphical analysis of the data and corresponding fits exhibit consistent behavior by material type between the two laboratories. Sources of measurement variability are addressed.Keywords
This publication has 12 references indexed in Scilit:
- Fitting Linear Mixed-Effects Models Usinglme42015
- Analysis of Thermal-Conductivity Measurement Data from International Comparison of National LaboratoriesInternational Journal of Thermophysics, 2012
- Thermal conductivity measurements of insulating materials: results of a round-robin test between laboratories and manufacturersHigh Temperatures-High Pressures, 1997
- Report on Inter-Laboratory Comparison of Guarded Hot Plate MeasurementsJournal of Thermal Insulation and Building Envelopes, 1995
- Thermal imbalance error and effective area in guarded hot plates with anisotropic specimensReview of Scientific Instruments, 1987
- Circular and Square Edge Effect Study for Guarded-Hot-Plate and Heat-Flow-Meter ApparatusesJournal of Thermal Insulation, 1986
- Thermal imbalance errors and effective area in guarded hot platesReview of Scientific Instruments, 1986
- Present and Future Research on Guarded Hot Plate and Heat Flow Meter Apparatus1985
- Design Criteria for Guarded Hot Plate Apparatus1974
- Deviations from One-Dimensional Heat Flow in Guarded Hot-Plate MeasurementsReview of Scientific Instruments, 1957