Abstract
Objective: To determine the effect of differences in the distances and radial positions of the tips of Light‐Curing Units (LCU) on the irradiance values delivered. Materials and Methods: Four LCUs —Valo X, Emitter A, CuringPen , and Pinkwave were tested at different distances from the tip (0, 2, 4, 6, 8 mm) and radial displacements (2 and 4 mm East, West, North, South). The radiant power ( mW ) and emission spectrum ( mW /nm) were measured using an integrating sphere with a 4 mm aperture. Calibrated irradiance beam profiles were recorded across the tip using a laser beam analyzer. Data were analyzed with a two‐way ANOVA at a significance level of 5%. Results: Radial displacement significantly reduced the irradiance values delivered from all four LCUs ( p < 0.05). At 4 mm, the mean irradiance value was below 1000 mW /cm 2 for Emitter A (256 mW/cm 2 ), CuringPen (573 mW/cm 2 ), and Valo X (955 mW/cm 2 ), but not for the Pinkwave (1157 mW /cm 2 ). At the 2 and 4 mm distance from the tip, the Valo X delivered consistent mean irradiance values at both the center and radial positions (2 mm: 1120 mW /cm 2 and 1089 mW /cm 2 ; 4 mm: 979 mW /cm 2 and 945 mW /cm 2 , p > 0.05). The Valo X emitted the most homogeneous beam profile, irrespective of the distance. Conclusion: Variations of 2 or 4 mm in the radial tip position negatively affected irradiance values from the LCUs tested. Clinical Significance: The radial position and LCU choice can significantly affect the irradiance delivered.