Usefulness of a lead-acrylic shield for reducing lens dose of assistant in x-ray CT examination
- 1 June 2022
- journal article
- research article
- Published by IOP Publishing in Journal of Radiological Protection
- Vol. 42 (2) , 021529
Abstract
In computed tomography (CT) examinations, the usefulness of protective glasses for reducing lens exposure to assistants has been reported. The present study aimed to compare the dose reduction effect for assistants with lead-acrylic shields and protective glasses (0.07 mm Pb, 0.5 mm Pb) during CT examination. The air dose distribution in a CT examination room with and without a lead-acrylic shield was compared. It was found that the amount of scattered radiation was significantly reduced by installing a lead-acrylic shield at the CT gantry aperture. Moreover, the reduction rate of air kerma at the assistant’s lens was higher using the lead acrylic shield than with the protective glasses—95.7% during head holding and 76.1% during assisted ventilation.Keywords
This publication has 25 references indexed in Scilit:
- A New Design of a Lead-Acrylic Shield for Staff Dose Reduction in Radial and Femoral Access Coronary CatheterizationRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2015
- Evaluation of Radiation Exposure of Medical Staff During CT-Guided InterventionsJournal of the American College of Radiology, 2015
- Radiation exposure of the radiologist’s eye lens during CT-guided interventionsActa Radiologica, 2014
- Air kerma to Hp(3) conversion coefficients for a new cylinder phantom for photon reference radiation qualitiesRadiation Protection Dosimetry, 2012
- ICRP PUBLICATION 118: ICRP Statement on Tissue Reactions and Early and Late Effects of Radiation in Normal Tissues and Organs — Threshold Doses for Tissue Reactions in a Radiation Protection ContextAnnals of the ICRP, 2012
- ESTIMATION OF THE NUMBER OF CT PROCEDURES BASED ON A NATIONWIDE SURVEY IN JAPANHealth Physics, 2011
- Comparing Strategies for Operator Eye Protection in the Interventional Radiology SuiteJournal of Vascular and Interventional Radiology, 2010
- Occupational Radiation Protection in Interventional Radiology: A Joint Guideline of the Cardiovascular and Interventional Radiology Society of Europe and the Society of Interventional RadiologyJournal of Vascular and Interventional Radiology, 2010
- CT Fluoroscopy Shielding: Decreases in Scattered Radiation for the Patient and OperatorJournal of Vascular and Interventional Radiology, 2006
- Impact of a lead glass screen on scatter radiation to eyes and hands in interventional cardiologistsCatheterization and Cardiovascular Interventions, 2005