End-tidal versus manually-controlled low-flow anaesthesia
- 11 October 2013
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 28 (2) , 117-121
Abstract
During low-flow manually-controlled anaesthesia (MCA) the anaesthetist needs constantly adjust end-tidal oxygen (EtO2) and anaesthetic concentrations (EtAA) to assure an adequate and safe anaesthesia. Recently introduced anaesthetic machines can automatically maintain those variables at target values, avoiding the burden on the anaesthetist. End-tidal-controlled anaesthesia (EtCA) and MCA provided by the same anaesthetic machine under the same fresh gas flow were compared. Eighty patients were prospectively observed: in MCA group (n = 40) target end-tidal sevoflurane (1 %) and EtO2 concentrations (≥35 %) were manually controlled by the anaesthetist. In EtCA group (n = 40) the same anaesthetic machine with an additional end-tidal control feature was used to reach the same targets, rendering automatic the achievement and maintenance of those targets. Anaesthetic machine characteristics, amount of consumed gases, oxygen and sevoflurane efficiencies, and the amount of interventions by the anaesthetist were recorded. In EtCA group EtAA was achieved later (145 s) than in MCA (71 s) and remained controlled thereafter. Even though the target expired gas fractions were achieved faster in MCA, manual adjustments were required throughout anaesthesia for both oxygen and sevoflurane. In MCA patients the number of manual adjustments to stabilize EtAA and EtO2 were 137 and 107, respectively; no adjustment was required in EtCA. Low-flow anaesthesia delivered with an anaesthetic machine able to automatically control EtAA and EtO2 provided the same clinical stability and avoided the continuous manual adjustment of delivered sevoflurane and oxygen concentrations. Hence, the anaesthetist could dedicate more time to the patient and operating room activities.Keywords
This publication has 20 references indexed in Scilit:
- Assessing the clinical or pharmaco‐economical benefit of target controlled desflurane delivery in surgical patients using the Zeus ® anaesthesia machine*Anaesthesia, 2009
- Desflurane consumption during automated closed-circuit delivery is higher than when a conventional anesthesia machine is used with a simple vaporizer-O2-N2O fresh gas flow sequenceBMC Anesthesiology, 2008
- Changing Patterns in Anesthetic Fresh Gas Flow Rates Over 5 Years in a Teaching HospitalAnesthesia & Analgesia, 2008
- Low-flow anaesthesiaContinuing Education in Anaesthesia Critical Care & Pain, 2008
- Tutorial: Context-Sensitive Decrement Times for Inhaled AnestheticsAnesthesia & Analgesia, 2005
- Low-flow anesthesia: Theory, practice, technical preconditions, advantages, and foreign gas accumulationJournal of Anesthesia, 1999
- Context-Sensitive Half-Times and Other Decrement Times of Inhaled AnestheticsAnesthesia & Analgesia, 1997
- Context-Sensitive Half-Times and Other Decrement Times of Inhaled AnestheticsAnesthesia & Analgesia, 1997
- Low and minimal flow inhalational anaesthesiaCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1997
- Algorithm AS 62: A Generator for the Sampling Distribution of the Mann- Whitney U StatisticJournal of the Royal Statistical Society Series C: Applied Statistics, 1973