Optimization of Patient Management in the Gynecology Emergency Department Using Point-of-Care Beta hCG
Open Access
- 9 July 2022
- journal article
- research article
- Published by MDPI AG in Diagnostics
- Vol. 12 (7) , 1670
Abstract
Background: Point-of-care testing (POCT) provides shorter turn-around times and, in many cases, potentially improves medical decision making. The AQT90 FLEX® benchtop immunoanalyzer (Radiometer Medical ApS, Copenhagen, Denmark) allows for the determination of beta-human chorionic gonadotropin (βhCG) in 18 min. The main aim of this study was to evaluate the impact of measuring βhCG using the AQT90 analyzer in the gynecology emergency department (ED) compared to the standard practice of using central laboratory blood testing on the patient length of stay (LOS). Methods: The evaluation consisted of two parts. The first one, conducted in the central laboratory, focused on the analytical performances of the AQT βhCG assay. The second one, conducted in the ED, aimed at determining the impact of POCT βhCG implementation on the timeframe in which ED patients require βhCG assessment. Results: The within-lab imprecisions at the mean values of 17 and 287 IU/L were 2.7% and 3.7%, respectively. Using Deming regression (n = 60), the following equation was obtained in the central lab: AQT90 βhCG = 1.1 Roche βhCG—12.9 (r = 0.997). The implementation of POCT βhCG in the ED significantly reduced patient LOS (145 (90–212) min vs. 205 (155–265) with and without AQT90, respectively, p < 0.001). At the 2 IU/L decision level, a 99.7% agreement with the Roche assay was reported (kappa statistics, 0.99). Conclusions: We confirm that the analytical qualities of the AQT 90 were in line with those obtained in the central lab. The implementation of the POCT βhCG is associated with a shorter LOS in the ED due to the faster availability of the results and the faster decision-making possibilities.Keywords
Funding Information
- the Radiometer Medical ApS (Copenhagen, Denmark)
This publication has 33 references indexed in Scilit:
- Total hCG testsClinica Chimica Acta, 2011
- A systematic review of triage-related interventions to improve patient flow in emergency departmentsScandinavian Journal of Trauma, Resuscitation and Emergency Medicine, 2011
- Implementation of a new emergency medical communication centre organization in Finland - an evaluation, with performance indicatorsScandinavian Journal of Trauma, Resuscitation and Emergency Medicine, 2011
- The analytical specificity of human chorionic gonadotropin assays determined using WHO International Reference ReagentsClinica Chimica Acta, 2010
- Implementation of a Rapid Whole Blood D-Dimer Test in the Emergency Department of an Urban Academic Medical CenterAmerican Journal of Clinical Pathology, 2009
- Differences in Recognition of the 1st WHO International Reference Reagents for hCG-Related Isoforms by Diagnostic Immunoassays for Human Chorionic GonadotropinClinical Chemistry, 2009
- Background hCG in non-pregnant individuals: Need for more sensitive point-of-care and over-the-counter pregnancy testsClinical Biochemistry, 2009
- Diagnostic Considerations in the Measurement of Human Chorionic Gonadotropin in Aging WomenClinical Chemistry, 2005
- Two-center clinical evaluation of a new automated fluorometric immunoassay for the quantitative analysis of total βeta-human chorionic gonadotropinClinical Biochemistry, 2003
- INTERNATIONAL REFERENCE PREPARATION OF HUMAN CHORIONIC GONADOTROPHIN FOR IMMUNOASSAY: POTENCY ESTIMATES IN VARIOUS BIOASSAY AND PROTEIN BINDING ASSAY SYSTEMS; AND INTERNATIONAL REFERENCE PREPARATIONS OF THE α AND β SUBUNITS OF HUMAN CHORIONIC GONADOTROPHIN FOR IMMUNOASSAYJournal of Endocrinology, 1980