Counting photoelectrons in the response of a photomultiplier tube to single picosecond light pulses
- 1 August 2004
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 75 (8) , 2762-2765
Abstract
We demonstrate the capability of a commercial photomultiplier to produce distinguishable anodic-pulse charge values when 1, 2, or more photoelectrons leave the cathode within a time shorter than the pulse-response duration. We propose a method for precise reconstruction of the photoelectron statistics from the measured pulse-height spectra and discuss applications to the characterization of quantum states in the continuous-variable regime.Keywords
This publication has 8 references indexed in Scilit:
- Fiber-assisted detection with photon number resolutionOptics Letters, 2003
- Conditional Preparation of a Quantum State in the Continuous Variable Regime: Generation of a sub-Poissonian State from Twin BeamsPhysical Review Letters, 2003
- Photon-number resolution using time-multiplexed single-photon detectorsPhysical Review A, 2003
- Silicon photomultiplier and its possible applicationsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- Verification of parametric sources of entanglementJournal of Optics B: Quantum and Semiclassical Optics, 2002
- A robust verification of the quantum nature of lightPhysics Letters A, 2001
- Multiphoton detection using visible light photon counterApplied Physics Letters, 1999
- Why people like the Hybrid PhotoDiodeNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997