Abstract
Compton imaging is an attractive tool for range verification and dose estimation in proton therapy. To investigate the application of CdZnTe based cameras for this task, we apply the uniform Cramér-Rao bound to dual-plane based designs with various interplane spacings and evaluate their bias-variance tradeoff. The investigation focuses on the 718 keV and 4.4 MeV prompt gammas emitted from proton interactions with 12C, and focuses on a detector geometry having two planes of 3×3 CdZnTe crystals with a volume of 2×2×1.5 cm3. When considering only interplane events, the improvement in minimum variance plateaus at 8 cm. However, when considering both intra-and inter-plane events with an efficiency factor, the optimum spacing is calculated to be around 4 cm as larger spacings degrade the performance with poorer efficiency and a lower interplane to intraplane fraction. In addition, the study uses the modified uniform Cramér-Rao bound based on a simulated distribution of prompt gamma rays expected from proton irradiation. The results show that the optimum spacing may be between 24 cm, depending on the specific bias-gradient norm.