Cramér–Rao Bound Evaluations of Compton Imager Designs for Proton Beam Range Verification
- 28 September 2021
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Radiation and Plasma Medical Sciences
- Vol. 6 (6) , 731-739
Abstract
Compton imaging is an attractive tool for range verification and dose estimation in proton therapy. To investigate the application of CdZnTe based cameras for this task, we apply the uniform Cramér-Rao bound to dual-plane based designs with various interplane spacings and evaluate their bias-variance tradeoff. The investigation focuses on the 718 keV and 4.4 MeV prompt gammas emitted from proton interactions with 12C, and focuses on a detector geometry having two planes of 3×3 CdZnTe crystals with a volume of 2×2×1.5 cm3. When considering only interplane events, the improvement in minimum variance plateaus at 8 cm. However, when considering both intra-and inter-plane events with an efficiency factor, the optimum spacing is calculated to be around 4 cm as larger spacings degrade the performance with poorer efficiency and a lower interplane to intraplane fraction. In addition, the study uses the modified uniform Cramér-Rao bound based on a simulated distribution of prompt gamma rays expected from proton irradiation. The results show that the optimum spacing may be between 24 cm, depending on the specific bias-gradient norm.Keywords
Funding Information
- Department of Energy NA-22 office and the Consortium for Monitoring, Technology, and Verification (DE-NA0003920)
This publication has 47 references indexed in Scilit:
- Imaging of prompt gamma rays emitted during delivery of clinical proton beams with a Compton camera: feasibility studies for range verificationPhysics in Medicine & Biology, 2015
- Noise evaluation of Compton camera imaging for proton therapyPhysics in Medicine & Biology, 2015
- Proton range verification through prompt gamma-ray spectroscopyPhysics in Medicine & Biology, 2014
- Range assessment in particle therapy based on promptγ-ray timing measurementsPhysics in Medicine & Biology, 2014
- Detecting prompt gamma emission during proton therapy: the effects of detector size and distance from the patientPhysics in Medicine & Biology, 2014
- Optimizing a three-stage Compton camera for measuring prompt gamma rays emitted during proton radiotherapyPhysics in Medicine & Biology, 2010
- Design Guidelines for a Double Scattering Compton Camera for Prompt-$\gamma$ Imaging During Ion Beam Therapy: A Monte Carlo Simulation StudyIEEE Transactions on Nuclear Science, 2010
- Uncertainty analysis of a Compton camera imaging system for radiation therapy dose reconstructionMedical Physics, 2010
- Statistical performance evaluation and comparison of a Compton medical imaging system and a collimated Anger camera for higher energy photon imagingPhysics in Medicine & Biology, 2008
- Feasibility study of using hybrid collimation for nuclear environmental imagingIEEE Transactions on Nuclear Science, 2003