Continuous, long-term hemodynamic monitoring is crucial for enhancing the prevention and diagnosis of cardiovascular diseases (CVDs). This study introduces a system that extracts impedance cardiography (ICG), electrocardiography (ECG), and respiratory (RESP) signals from thoracic electrical signals using only four electrodes. A wireless, portable hemodynamic monitoring system has been developed, capable of acquiring multimodal physiological signals, including ICG, ECG, and photoplethysmography (PPG), among others. A signal processing approach was developed, combining the timefrequency characteristics and temporal correlations of the signals. This method includes signal denoising and locating characteristic points, enabling the real-time calculation of hemodynamic parameters such as cardiac output (CO). Furthermore, an artificial neural network is constructed based on various features extracted from the multimodal physiological signals to estimate blood pressure (BP). The device measures $85\times 48\times 18$ mm, weighs 76 g, and consumes 260 mW of power. It can operate continuously for at least 24 h. When compared to a commercial bedside hemodynamic monitoring device, the consistency of the key measured hemodynamic parameters is relatively high. The system shows promising applications in diagnosing CVDs and managing chronic cardiovascular conditions.Keywords
Funding Information
- National Natural Science Foundation of China (81971700)
- Graduate Research Innovation Project in Chongqing, China, (CYS23135)
This publication has 43 references indexed in Scilit:
- A review: Blood pressure monitoring based on PPG and circadian rhythmAPL Bioengineering, 2024
- Global Burden of Cardiovascular Diseases and Risks, 1990-2022JACC, 2023
- A two-electrode frequency-scan system for gesture recognitionMechatronics, 2023
- Signal processing techniques applied to impedance cardiography ICG signals – a reviewJournal of Medical Engineering & Technology, 2022
- Clinical Use of Impedance Cardiography for Hemodynamic Assessment of Early Cardiovascular Disease and Management of HypertensionHigh Blood Pressure & Cardiovascular Prevention, 2020
- Detection and Removal of Motion Artifacts in PPG SignalsMobile Networks and Applications, 2019
- Design of Noninvasive Hemodynamic Monitoring Equipment Using Impedance Cardiography2019
- Accuracy of Impedance Cardiography for Hemodynamic Assessment During Rest and Exercise in Wheelchair Rugby PlayersResearch Quarterly for Exercise and Sport, 2019
- Do mathematical model studies settle the controversy on the origin of cardiac synchronous trans-thoracic electrical impedance variations? A systematic reviewPhysiological Measurement, 2016
- The IMPACT shirt: textile integrated and portable impedance cardiographyPhysiological Measurement, 2014