Feasibility of using linearly polarized rotating birdcage transmitters and close-fitting receive arrays in MRI to reduce SAR in the vicinity of deep brain simulation implants
- 5 April 2016
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 77 (4) , 1701-1712
Abstract
Purpose MRI of patients with deep brain stimulation (DBS) implants is strictly limited due to safety concerns, including high levels of local specific absorption rate (SAR) of radiofrequency (RF) fields near the implant and related RF‐induced heating. This study demonstrates the feasibility of using a rotating linearly polarized birdcage transmitter and a 32‐channel close‐fit receive array to significantly reduce local SAR in MRI of DBS patients. Methods Electromagnetic simulations and phantom experiments were performed with generic DBS lead geometries and implantation paths. The technique was based on mechanically rotating a linear birdcage transmitter to align its zero electric‐field region with the implant while using a close‐fit receive array to significantly increase signal to noise ratio of the images. Results It was found that the zero electric‐field region of the transmitter is thick enough at 1.5 Tesla to encompass DBS lead trajectories with wire segments that were up to 30 degrees out of plane, as well as leads with looped segments. Moreover, SAR reduction was not sensitive to tissue properties, and insertion of a close‐fit 32‐channel receive array did not degrade the SAR reduction performance. Conclusion The ensemble of rotating linear birdcage and 32‐channel close‐fit receive array introduces a promising technology for future improvement of imaging in patients with DBS implants. Magn Reson Med 77:1701–1712, 2017. © 2016 International Society for Magnetic Resonance in MedicineKeywords
Funding Information
- National Institutes of Health (P41EB015896, R21EB016449)
This publication has 47 references indexed in Scilit:
- Therapeutic Subthalamic Nucleus Deep Brain Stimulation Reverses Cortico-Thalamic Coupling during Voluntary Movements in Parkinson's DiseasePLOS ONE, 2012
- Reduction of the radiofrequency heating of metallic devices using a dual‐drive birdcage coilMagnetic Resonance in Medicine, 2012
- A numerical postprocessing procedure for analyzing radio frequency MRI coilsConcepts in Magnetic Resonance Part A, 2011
- Network perspectives on the mechanisms of deep brain stimulationNeurobiology of Disease, 2010
- 96‐Channel receive‐only head coil for 3 Tesla: Design optimization and evaluationMagnetic Resonance in Medicine, 2009
- Deep Brain Stimulation of the Ventral Capsule/Ventral Striatum for Treatment-Resistant DepressionBiological Psychiatry, 2009
- MRI-based anatomical model of the human head for specific absorption rate mappingMedical & Biological Engineering & Computing, 2008
- Changes in cerebral activity pattern due to subthalamic nucleus or internal pallidum stimulation in Parkinson's diseaseAnnals of Neurology, 1997
- The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHzPhysics in Medicine & Biology, 1996
- The dielectric properties of biological tissues: I. Literature surveyPhysics in Medicine & Biology, 1996