Development of an Internal Calibration Algorithm for Ultrahigh-Resolution Mass Spectra of Dissolved Organic Matter
- 21 July 2022
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 94 (30) , 10589-10594
Abstract
In order to obtain a spectrum with high mass accuracy, an internal calibration of Fourier transform ion cyclotron resonance mass spectrometry (FTICR-MS) is inevitable. This in turn is critical for subsequent data processing and is generally performed using the commercial instrument software DataAnalysis in the benchmark calibration mode. However, no methodological study has systemically addressed the automated internal calibration of FTICR-MS spectra for dissolved organic matter (DOM) from different sources such as terrestrial and aquatic environments. In this study, a new piecewise algorithm, FTMSCalibrate, was developed to automatically calibrate FTICR-MS spectra in both positive and negative ion modes. FTMSCalibrate was found to reproduce 91.7% ± 4.4% (referred to as the true positive ratio) of the chemical formulas obtained by calibration using manual DataAnalysis. In addition to significantly reducing the mass error, FTMSCalibrate is more accurate in terms of the molecular formula assignment for low m/z peaks than Formularity and MFAssignR. FTMSCalibrate was compatible with deprotonated ions for FTICR-MS spectra in the negative ion mode as well as protonated and adduct ions, including Na- and K-adducts, for FTICR-MS spectra in the positive ion mode. These results suggest that FTMSCalibrate publicly available herein is a robust alternative for the internal calibration of FTICR-MS spectra during postdata processing and will facilitate DOM analysis by FTICR-MS.Keywords
Funding Information
- Japan Society for the Promotion of Science (17H04588, 19H02271)
- National Natural Science Foundation of China (42107484)
- China University of Geosciences (122-G1323522145)
This publication has 18 references indexed in Scilit:
- Comparative study of organic matter chemical characterization using negative and positive mode electrospray ionization ultrahigh-resolution mass spectrometryAnalytical and Bioanalytical Chemistry, 2016
- Iterative Method for Mass Spectra Recalibration via Empirical Estimation of the Mass Calibration Function for Fourier Transform Mass Spectrometry-Based PetroleomicsAnalytical Chemistry, 2013
- Parts-Per-Billion Fourier Transform Ion Cyclotron Resonance Mass Measurement Accuracy with a “Walking” Calibration EquationAnalytical Chemistry, 2011
- Naturally present fatty acids as internal calibrants for Fourier transform mass spectra of dissolved organic matterLimnology and Oceanography: Methods, 2008
- Fundamentals of Molecular Formula Assignment to Ultrahigh Resolution Mass Data of Natural Organic MatterAnalytical Chemistry, 2007
- Sub part-per-million mass accuracy by using stepwise-external calibration in fourier transform ion cyclotron resonance mass spectrometryJournal of the American Society for Mass Spectrometry, 2006
- Characterization of a major refractory component of marine dissolved organic matterGeochimica et Cosmochimica Acta, 2006
- Statistical Evaluation of Internal and External Mass Calibration Laws Utilized in Fourier Transform Ion Cyclotron Resonance Mass SpectrometryAnalytical Chemistry, 2005
- Accurate mass measurements by Fourier transform mass spectrometryMass Spectrometry Reviews, 2004
- Routine Part-per-Million Mass Accuracy for High- Mass Ions: Space-Charge Effects in MALDI FT-ICRAnalytical Chemistry, 1998