Abstract
Modifying surface topography to prevent surface-induced thrombosis in cardiovascular implants allows endothelialization, which is the natural thrombo-resistance of the blood-contacting surfaces and is deemed to be the only long-term hemocompatible material. We have adapted a simulation framework to predict platelet deposition on the modified surface and developed an optimization strategy to promote endothelial retention and limit platelet deposition. Under supraphysiological bulk shear stress, a maximum of 79% linear coverage was achieved. This study concludes that the addition of microtrenches promotes endothelial retention and can be improved through optimal selection of geometric parameters.