No correlation between radiolucency and biomechanical stability of keeled and pegged glenoid components
Authors:
- Andreas VossAVAndreas Voss
- Knut BeitzelKBKnut Beitzel
- Elifho ObopilweEOElifho Obopilwe
- Stefan BuchmannSBStefan Buchmann
- John ApostolakosJAJohn Apostolakos
- Jessica Di VenereJVJessica Di Venere
- Michael NowakMNMichael Nowak
- Mark P. CoteMCMark P. Cote
- Anthony A. RomeoARAnthony A. Romeo
- Augustus D. MazzoccaAMAugustus D. Mazzocca
Open Access
- 25 May 2017
- journal article
- research article
- Published by Springer Nature in BMC Musculoskeletal Disorders
- Vol. 18 (1) , 1-11
Abstract
The purpose of this study was to examine biomechanical properties and the degree of radiolucency of two cemented basic glenoid designs for total shoulder arthroplasty. Our hypothesis was that a component with increased micro-motion in the laboratory at time zero would also exhibit a greater amount of radiolucency in patients at a minimum of 2 years post total shoulder arthroplasty. Thirty cadaveric shoulders were divided into 2 groups (keel vs. peg). The glenoid components were first loaded with a single axial eccentric force of 196 N in all orientations and then with a transversal load of 49 N to simulate in vivo loads with abduction. Displacement of the glenoid component was determined with four different linear variable-differential transducers. In the second phase, 56 antero-posterior x-rays of 52 patients with either the same keeled (n = 24) or pegged (n = 32) glenoid component with a minimum of 24 months follow-up were evaluated for radiolucency. Biomechanically the pegged glenoid showed a significant increase in micro-motion during eccentric axial loading as well as during combined loading in the anterior, posterior, and inferior position as compared to the keeled glenoid (p < 0.05). In contrast all results were significant with greater radiolucency for the keeled glenoid component (p = 0.001). While the pegged component exhibited a greater amount of micro-motion during biomechanical testing, radiolucency was greater in patients with a keeled component. These findings provide support for both components from different perspectives and highlight the need for well-constructed studies to determine whether glenoid design has an effect on clinical outcome, because influences are multifactorial and biomechanical forces may not recreate forces seen in vivo.Keywords
Funding Information
- Arthrex (US)
- German Research Foundation (DFG) and the Technische Universität München
This publication has 26 references indexed in Scilit:
- Radiographic survival in total shoulder arthroplastyJournal of Shoulder and Elbow Surgery, 2013
- Glenoid morphology after reaming in computer-simulated total shoulder arthroplastyJournal of Shoulder and Elbow Surgery, 2013
- Pegged versus keeled glenoid components in total shoulder arthroplastyJournal of Shoulder and Elbow Surgery, 2010
- Radiographic comparison of pegged and keeled glenoid components using modern cementing techniques: A prospective randomized studyJournal of Shoulder and Elbow Surgery, 2010
- Survival of the glenoid component in shoulder arthroplastyJournal of Shoulder and Elbow Surgery, 2009
- Glenoid Component Failure in Total Shoulder ArthroplastyJournal of Bone and Joint Surgery, 2008
- Complications of Total Shoulder ArthroplastyJournal of Bone and Joint Surgery, 2006
- Complications of total shoulder arthroplasty: Are they fewer or different?Journal of Shoulder and Elbow Surgery, 2006
- Radiographic comparison of pegged and keeled glenoid componentsJournal of Shoulder and Elbow Surgery, 2005
- Finite element modeling of the glenoid component: Effect of design parameters on stress distributionJournal of Shoulder and Elbow Surgery, 1992