Operating a triple stack microchannel plate-phosphor assembly for single particle counting in the 12–300K temperature range
- 1 November 2007
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 78 (11) , 113301
Abstract
An assembly consisting of a stack of three microchannel plates (MCPs) and a phosphor screen anode has been operated over the temperature range from . We report on measurements at (using an alpha source) and with dark counts only . Without any particle source, the MCP bias current decreased by a factor of when the temperature was lowered from . Using the alpha source, and a photomultiplier tube (PMT) to monitor the phosphor screen anode, we first observed an increase in the decay time of the phosphor from when the temperature was decreased from while the decay time then decreased and reached a value of at . The pulse height distribution from the PMT was measured between 300 and and shows a spectrum typical for a MCP phosphor setup at and but is strongly degraded for intermediate temperatures. We conclude that the present MCP-phosphor detector assembly is well suited for position-sensitive particle counting operation at temperatures down to at least even for count rates beyond . This result is crucial and an important part of ongoing developments of new instrumentation for investigations of, e.g., interactions involving complex molecular ions with internal quantum state control.
Keywords
This publication has 9 references indexed in Scilit:
- Study of indicators for the development of fluorescence based optical fiber temperature sensorsSensors and Actuators B: Chemical, 2006
- DESIREE — A Double Electrostatic Storage Ring for Merged-Beam ExperimentsAIP Conference Proceedings, 2006
- A multi-channel optical fibre thermometerMeasurement, 2005
- Physics with colder molecular ions: The Heidelberg Cryogenic Storage Ring CSRJournal of Physics: Conference Series, 2005
- Dissociative recombination and excitation of O2+: Cross sections, product yields and implications for studies of ionospheric airglowsThe Journal of Chemical Physics, 2001
- Microchannel plate resistance at cryogenic temperaturesNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2000
- Phosphor thermometry at cryogenic temperaturesReview of Scientific Instruments, 1997
- Optical fibre temperature sensor in the cryogenic rangeOptics & Laser Technology, 1996
- The performance of a microchannel plate at cryogenic temperatures and in high magnetic fields, and the detection efficiency for low energy positive hydrogen ionsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1992