Operating a triple stack microchannel plate-phosphor assembly for single particle counting in the 12–300K temperature range

Authors:

Abstract
An assembly consisting of a stack of three microchannel plates (MCPs) and a phosphor screen anode has been operated over the temperature range from 300to12K . We report on measurements at 6.4kHz (using an alpha source) and with dark counts only (15Hz) . Without any particle source, the MCP bias current decreased by a factor of 2.1×103 when the temperature was lowered from 300to12K . Using the alpha source, and a photomultiplier tube (PMT) to monitor the phosphor screen anode, we first observed an increase in the decay time of the phosphor from 12to45μs when the temperature was decreased from 300to100K while the decay time then decreased and reached a value of 5μs at 12K . The pulse height distribution from the PMT was measured between 300 and 12K and shows a spectrum typical for a MCP phosphor setup at 300K and 12K but is strongly degraded for intermediate temperatures. We conclude that the present MCP-phosphor detector assembly is well suited for position-sensitive particle counting operation at temperatures down to at least 12K even for count rates beyond 6kHz . This result is crucial and an important part of ongoing developments of new instrumentation for investigations of, e.g., interactions involving complex molecular ions with internal quantum state control.

This publication has 9 references indexed in Scilit: