Abstract
We have compared the stability of two distinct atomic frequency standards. One is based on an atomic cold cloud of 133Cs atoms in free expansion, interrogated by a Rabi resonance. The other standard is based on a thermal Cs beam interrogated by the separated oscillatory field method (Ramsey’s method). The atomic expanding cloud standard shows better stability and technical advances based on its simplicity.Keywords
This publication has 6 references indexed in Scilit:
- Atomic clocks based on coherent population trapping: a reviewApplied Physics B, 2005
- Microfabricated atomic frequency referencesMetrologia, 2005
- Characterization of the Brazilian Cs atomic-frequency standard: evaluation of major shiftsMetrologia, 2002
- Construction and evaluation of the first Brazilian atomic clockIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2000
- A cesium fountain frequency standard: preliminary resultsIEEE Transactions on Instrumentation and Measurement, 1995
- Observation of the cesium clock transition in laser-cooled atomsOptics Letters, 1989