Atraumatic Fracture of the BIOLOX delta Ceramic Liner in Well-Fixed Total Hip Implants
Authors:
- 1 November 2018
- journal article
- research article
- Published by SLACK, Inc. in Orthopedics
- Vol. 41 (6) , E880-E883
Abstract
A newer alumina matrix composite ceramic (BIOLOX delta; CeramTec, Plochingen, Germany) was developed to improve resistance to wear while reducing the risk for fractures with alumina-on-alumina design. This article describes a case of fracture of the BIOLOX delta ceramic liner without history of trauma in well-fixed total hip implants. A 61-year-old woman underwent total hip replacement via the posterior approach 20 months earlier because of osteonecrosis of the femoral head. A cementless cup and a stem with a 52/36-mm BIOLOX delta couple bearing were used. The authors found signs of stripe wear on the inner surface of the ceramic liner, leading them to believe the possible cause of fracture in this patient was edge loading due to joint laxity. The authors performed revision surgery with the use of another ceramic couple-bearing implant and converted the head component from short to long (+ 4 mm). More cases should be collected to verify the mechanism of the fracture of the BIOLOX delta ceramic liner with a well-fixed cup.Keywords
This publication has 13 references indexed in Scilit:
- Mid-term results of the BIOLOX delta ceramic-on-ceramic total hip arthroplastyThe Bone & Joint Journal, 2017
- The Current Concepts of Total Hip ArthroplastyHip & Pelvis, 2016
- Wear of a Composite Ceramic Head Caused by Liner FractureOrthopedics, 2014
- Fracture of the Alumina-bearing Couple Delta Ceramic LinerOrthopedics, 2012
- THA With Delta Ceramic on Ceramic: Results of a Multicenter Investigational Device Exemption TrialClinical Orthopaedics and Related Research, 2010
- Ceramic-on-Ceramic Total Hip ArthroplastyThe Journal of Arthroplasty, 2009
- Chromosomal Aberrations in the Peripheral Blood of Patients with Metal-on-Metal Hip BearingsJournal of Bone and Joint Surgery, 2008
- Lining Up the Liner: 2 Case Reports of Early Ceramic Liner FragmentationThe Journal of Arthroplasty, 2007
- Ceramic Liner Fracture after Cementless Alumina-on-Alumina Total Hip ArthroplastyClinical Orthopaedics and Related Research, 2007
- Alumina-on-Alumina Total Hip ArthroplastyJournal of Bone and Joint Surgery, 2002