Optically Stimulated Luminescent Dosimetry for High Dose Rate Brachytherapy
Open Access
- 1 January 2012
- journal article
- research article
- Published by Frontiers Media SA in Frontiers in Oncology
- Vol. 2, 30763
Abstract
Purpose: The objective was to determine whether optically stimulated luminescent dosimeters (OSLDs) were appropriate for in vivo measurements in high dose rate brachytherapy. In order to make this distinction, three dosimetric characteristics were tested: dose linearity, dose rate dependence, and angular dependence. The Landauer nanoDot™ OSLDs were chosen due to their popularity and their availability commercially. Methods: To test the dose linearity, each OSLD was placed at a constant location and the dwell time was varied. Next, in order to test the dose rate dependence, each OSLD was placed at different OLSD-to-source distances and the dwell time was held constant. A curved geometry was created using a circular Accuboost® applicator in order to test angular dependence. Results: The OSLD response remained linear for high doses and was independent of dose rate. For doses up to 600 cGy, the linear coefficient of determination was 0.9988 with a response of 725 counts per cGy. The angular dependence was significant only in “edge-on” scenarios. Conclusion: OSLDs are conveniently read out using commercially available readers. OSLDs can be re-read and serve as a permanent record for clinical records or be annealed using conventional fluorescent light. Lastly, OSLDs are produced commercially for $5 each. Due to these convenient features, in conjunction with the dosimetric performance, OSLDs should be considered a clinically feasible and attractive tool for in vivo HDR brachytherapy measurements.Keywords
This publication has 25 references indexed in Scilit:
- A dosimetric uncertainty analysis for photon‐emitting brachytherapy sources: Report of AAPM Task Group No. 138 and GEC‐ESTROMedical Physics, 2011
- Dosimetric characterization of round HDR AccuBoost applicators for breast brachytherapyMedical Physics, 2009
- Task Group 142 report: Quality assurance of medical acceleratorsa)Medical Physics, 2009
- Optically stimulated luminescence (OSL) dosimetry in medicinePhysics in Medicine & Biology, 2008
- Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurementsMedical Physics, 2007
- Optically stimulated luminescence (OSL) of carbon‐doped aluminum oxide for film dosimetry in radiotherapyMedical Physics, 2007
- High-precision dosimetry for radiotherapy using the optically stimulated luminescence technique and thin Al2O3:C dosimetersPhysics in Medicine & Biology, 2005
- Monte Carlo‐aided dosimetry of a new high dose‐rate brachytherapy sourceMedical Physics, 1998
- Code of practice for brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 56Medical Physics, 1997
- Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43Medical Physics, 1995