GenVFNet: Generating Visual Field From Optical Coherence Tomography Angiography by Conditional Generative Adversarial Networks
Open Access
- 20 September 2024
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Access
- Vol. 12, 145845-145856
Abstract
Glaucoma investigations play an important role in glaucoma management. This study aimed to generate a visual field pattern deviation map (GenVF) from optical coherence tomography angiography (OCTA) using conditional generative adversarial networks (cGAN), namely GenVFNet. cGAN consisted of two components: generator and discriminator. The generator consisted of encoder and decoder blocks and the discriminator was the classification learner. The training data was Grad-CAM of OCTA (GradOCTA) of glaucomatous eyes. Only eyes with a visual field (VF) compatible with optic disc photograph (DP) and/or optical coherence tomography (OCT) were used. 100 eyes (90.9%) were used for training and the remaining 10 eyes (9.1%) for testing. To conform with the five severity levels in clinical diagnosis, we quantized the generated VF (GenRawVF) from GenVFNet into five levels and named the quantized image GenVF. In the experiment, GenRawVFs were compared with actual pattern deviation images (RealRawVF) using the structural similarity index measure (SSIM), normalized root mean square error (NRMSE), Fréchet inception distance (FID), and confusion matrix. The average ± 95%CI SSIM, NRMSE, and FID were 0.87 ± 0.02, 0.10 ± 0.6, and 7.54 ± 3.47, respectively. GenVFNet was able to construct GenVF, especially in the superior hemifield. In addition, GenVFNet was tested with VFs of eyes whose DP and/or OCT did not correlate with the available VF. According to the expert’s opinion, 81 eyes (65.8%) of GenVF were correlated with DP and/or OCT. This study demonstrated the potential of GenVFNet to construct a VF pattern deviation from the OCTA that would be a benefit for patients who cannot perform a reliable VF.Keywords
This publication has 44 references indexed in Scilit:
- Deep Learning Image Analysis of Optical Coherence Tomography Angiography Measured Vessel Density Improves Classification of Healthy and Glaucoma EyesAmerican Journal of Ophthalmology, 2022
- Predicting Visual Fields From Optical Coherence Tomography via an Ensemble of Deep Representation LearnersAmerican Journal of Ophthalmology, 2022
- Deep learning can generate traditional retinal fundus photographs using ultra-widefield images via generative adversarial networksComputer Methods and Programs in Biomedicine, 2020
- Estimating Global Visual Field Indices in Glaucoma by Combining Macula and Optic Disc OCT Scans Using 3-Dimensional Convolutional Neural NetworksOphthalmology Glaucoma, 2020
- Sectorwise Visual Field Simulation Using Optical Coherence Tomographic Angiography Nerve Fiber Layer Plexus Measurements in GlaucomaAmerican Journal of Ophthalmology, 2020
- Parapapillary Deep-Layer Microvasculature Dropout and Visual Field Progression in GlaucomaAmerican Journal of Ophthalmology, 2019
- Optical Coherence Tomography Angiography of the Peripapillary Retina in GlaucomaJAMA Ophthalmology, 2015
- Global Prevalence of Glaucoma and Projections of Glaucoma Burden through 2040Ophthalmology, 2014
- Confirmation of Visual Field Abnormalities in the Ocular Hypertension Treatment StudyArchives of Ophthalmology (1950), 2000
- Test-Retest Variability in Glaucomatous Visual FieldsAmerican Journal of Ophthalmology, 1989