A real‐time in vivo dosimetric verification method for high‐dose rate intracavitary brachytherapy of nasopharyngeal carcinoma
- 17 October 2012
- journal article
- radiation therapy-physics
- Published by Wiley in Medical Physics
- Vol. 39 (11) , 6757-6763
Abstract
Purpose: A real‐timein vivo dosimetric verification method using metal‐oxide‐semiconductor field effect transistor (MOSFET) dosimeters has been developed for patient dosimetry in high‐dose rate (HDR) intracavitary brachytherapy of nasopharyngeal carcinoma (NPC).Methods: The necessary calibration and correction factors for MOSFET measurements in192Iridium source were determined in a water phantom. With the detector placed inside a custom‐made nasopharyngeal applicator, the actual dose delivered to the tumor was measured in vivo and compared to the calculated values using a commercial brachytherapy planning system.Results: Five MOSFETs were independently calibrated with the HDR source, yielding calibration factors of 0.48 ± 0.007 cGy/mV. The maximum sensitivity variation was no more than 7% in the clinically relevant distance range of 1–5 cm from the source. A total of 70in vivo measurements in 11 NPC patients demonstrated good agreement with the treatment planning. The mean differences between the planned and the actually delivered dose within a single treatment fraction were −0.1% ± 3.8% and −0.1% ± 3.7%, respectively, for right and left side assessments. The maximum dose deviation was less than 8.5%.Conclusions: In vivo measurement using the real‐time MOSFET dosimetry system is possible to evaluate the actual dose to the tumor received by the patient during a treatment fraction and thus can offer another line of security to detect and prevent large errors.Keywords
Funding Information
- NNSFC (30900324)
This publication has 29 references indexed in Scilit:
- The effect of rectal heterogeneity on wall dose in high dose rate brachytherapyMedical Physics, 2008
- The Use of TLD and Gafchromic Film to Assure Submillimeter Accuracy for Image-Guided RadiosurgeryMedical Dosimetry, 2008
- A methodology for TLD postal dosimetry audit of high-energy radiotherapy photon beams in non-reference conditionsRadiotherapy and Oncology, 2007
- Verification of the plan dosimetry for high dose rate brachytherapy using metal–oxide–semiconductor field effect transistor detectorsMedical Physics, 2007
- Customized conformal high-dose-rate brachytherapy boost for limited-volume nasopharyngeal cancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2005
- Low-dose measurement with a MOSFET in high-energy radiotherapy applicationsRadiation Measurements, 2005
- Dose rate and SDD dependence of commercially available diode detectorsMedical Physics, 2004
- Feasibility study of online high-spatial-resolution MOSFET dosimetry in static and pulsed x-ray radiation fieldsIEEE Transactions on Nuclear Science, 2001
- Improved spatial resolution by MOSFET dosimetry of an x‐ray microbeamMedical Physics, 2000
- Evaluation of a dual bias dual metal oxide‐silicon semiconductor field effect transistor detector as radiation dosimeterMedical Physics, 1994