Abstract
Purpose: A real‐timein vivo dosimetric verification method using metal‐oxide‐semiconductor field effect transistor (MOSFET) dosimeters has been developed for patient dosimetry in high‐dose rate (HDR) intracavitary brachytherapy of nasopharyngeal carcinoma (NPC).Methods: The necessary calibration and correction factors for MOSFET measurements in192Iridium source were determined in a water phantom. With the detector placed inside a custom‐made nasopharyngeal applicator, the actual dose delivered to the tumor was measured in vivo and compared to the calculated values using a commercial brachytherapy planning system.Results: Five MOSFETs were independently calibrated with the HDR source, yielding calibration factors of 0.48 ± 0.007 cGy/mV. The maximum sensitivity variation was no more than 7% in the clinically relevant distance range of 1–5 cm from the source. A total of 70in vivo measurements in 11 NPC patients demonstrated good agreement with the treatment planning. The mean differences between the planned and the actually delivered dose within a single treatment fraction were −0.1% ± 3.8% and −0.1% ± 3.7%, respectively, for right and left side assessments. The maximum dose deviation was less than 8.5%.Conclusions: In vivo measurement using the real‐time MOSFET dosimetry system is possible to evaluate the actual dose to the tumor received by the patient during a treatment fraction and thus can offer another line of security to detect and prevent large errors.