Influence of Calibration Method on Distortion-Product Otoacoustic Emission Measurements: I. Test Performance
- 1 August 2010
- journal article
- research article
- Published by Wolters Kluwer Health in Ear & Hearing
- Vol. 31 (4) , 533-545
Abstract
Objective: Calibration errors in distortion-product otoacoustic emission (DPOAE) measurements because of standing waves cause unpredictable changes in stimulus and DPOAE response level. The purpose of this study was to assess the extent to which these errors affect DPOAE test performance. Standard calibration procedures use sound pressure level (SPL) to determine specified levels. Forward pressure level (FPL) is an alternate calibration method that is less susceptible to standing waves. However, FPL derivation requires prior cavity measurements, which have associated variability. In an attempt to address this variability, four FPL methods were compared with SPL: a reference calibration derived from 25 measurements before all data collection and a daily calibration measurement, both of which were made at body and room temperature. Design: Data were collected from 52 normal-hearing and 103 hearing-impaired subjects. DPOAEs were measured for f2 frequencies ranging from 2 to 8 kHz in half-octave steps, with L2 ranging from −20 to 70 dB SPL (5-dB steps). At each f2, DPOAEs were measured in five calibration conditions: SPL, daily FPL at body temperature (daily body), daily FPL at room temperature (daily room), reference FPL at body temperature (ref body), and reference FPL at room temperature (ref room). Data were used to construct receiver operating characteristic (ROC) curves for each f2, calibration method, and L2. From these curves, areas under the ROC curve (AROC) were estimated. Results: The results of this study are summarized by the following observations: (1) DPOAE test performance was sensitive to stimulus level, regardless of calibration method, with the best test performance observed for moderate stimulus level conditions. (2) An effect of frequency was observed for all calibration methods, with the best test performance at 6 kHz and the worst performance at 8 kHz. (3) At clinically applicable stimulus levels, little difference in test performance among calibration methods was noted across frequencies, except at 8 kHz. At 8 kHz, FPL-based calibration methods provided superior performance compared with the standard SPL calibration. (4) A difference between FPL calibration methods was observed at 8 kHz, with the best test performance occurring for daily calibrations at body temperature. Conclusions: With the exception of 8 kHz, there was little difference in test performance across calibration methods. At 8 kHz, AROCs and specificities for fixed sensitivities indicate that FPL-based calibration methods provide superior performance compared with the standard SPL calibration for clinically relevant levels. Temperature may have an impact on FPL calculations relative to DPOAE test performance. Although the differences in AROC among calibration procedures were not statistically significant, the present results indicate that standing wave errors may impact DPOAE test performance and can be reduced by using FPL, although the largest effects were restricted to 8 kHz.Keywords
This publication has 41 references indexed in Scilit:
- Influence of in situ, sound-level calibration on distortion-product otoacoustic emission variabilityThe Journal of the Acoustical Society of America, 2008
- Distortion-product otoacoustic emission: Early detection in deferoxamine induced ototoxicityAuris Nasus Larynx, 2008
- Distortion product otoacoustic emissions: Cochlear-source contributions and clinical test performanceThe Journal of the Acoustical Society of America, 2007
- Third Place—Resident Clinical Science Award 1998: Effects of Cis‐Platinum Chemotherapy on Otoacoustic Emissions: The Development of an Objective Screening ProtocolOtolaryngology–Head and Neck Surgery, 1999
- Effects of ear-canal standing waves on measurements of distortion-product otoacoustic emissionsThe Journal of the Acoustical Society of America, 1995
- Dependence of distortion-product otoacoustic emissions on primary levels in normal and impaired ears. I. Effects of decreasing L2 below L1The Journal of the Acoustical Society of America, 1995
- The spatial distribution of sound pressure within scaled replicas of the human ear canalThe Journal of the Acoustical Society of America, 1985
- Acoustical wave propagation in cylindrical ducts: Transmission line parameter approximations for isothermal and nonisothermal boundary conditionsThe Journal of the Acoustical Society of America, 1984
- Measurement of the acoustic input immittance of the human earThe Journal of the Acoustical Society of America, 1981
- The area above the ordinal dominance graph and the area below the receiver operating characteristic graphJournal of Mathematical Psychology, 1975