Comprehensive dose survey of breast screening in Ireland
Authors:
- 18 November 2010
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 145 (1) , 52-60
Abstract
The primary purpose of this study was to evaluate the impact of digital mammography screening on breast dose by analysing the results of a patient dose survey of the Irish breast screening programme. Results from the survey were used to determine a dose reference level for the screening programme. Approximately, 100 examinations were acquired for each of the digital mammography systems operational in the screening programme. Each examination consisted of two standard views of each breast. The mean glandular dose for each acquired image was calculated. The dose reference level was established by calculating the 95th percentile of the average mean glandular dose for the average compressed breast thickness of the mediolateral oblique views. The overall average mean glandular dose per examination was 2.72±0.04 mGy. The average compressed breast thickness was 61.4±0.03 mm. The average compression force was 109±7 N. A dose reference level value of 1.75 mGy was established for the screening programme. The results of this clinical dose survey provide a valuable indication of the dose performance of modern full field digital mammographic imaging systems. The results demonstrate clearly the dose benefits of digital mammography. The dose benefit of digital screening was further demonstrated by the establishment of a comparatively lower diagnostic reference level for the screening programme. The comparison of the dose performance of individual X-ray systems with the diagnostic reference level highlights the need for more optimisation within the service.Keywords
This publication has 9 references indexed in Scilit:
- Comparison of Acquisition Parameters and Breast Dose in Digital Mammography and Screen-Film Mammography in the American College of Radiology Imaging Network Digital Mammographic Imaging Screening TrialAmerican Journal of Roentgenology, 2010
- Further factors for the estimation of mean glandular dose using the United Kingdom, European and IAEA breast dosimetry protocolsPhysics in Medicine & Biology, 2009
- Results of a European dose survey for mammographyRadiation Protection Dosimetry, 2008
- Radiation doses received in the UK Breast Screening Programme in 2001 and 2002British Journal of Radiology, 2005
- Patient dose in full-field digital mammography: an Italian surveyEuropean Radiology, 2004
- Dose reduction in full-field digital mammography: an anthropomorphic breast phantom studyBritish Journal of Radiology, 2003
- Optimization of technique factors for a silicon diode array full‐field digital mammography system and comparison to screen‐film mammography with matched average glandular doseMedical Physics, 2003
- Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocolPhysics in Medicine & Biology, 2000
- Monte-Carlo calculation of conversion factors for the estimation of mean glandular breast dosePhysics in Medicine & Biology, 1990