Abstract
Purpose: The electron beams for total skin electron therapy (TSET) are often degraded by a scatter plate in addition to extended distances. For electron dosimetry, both the AAPM TG‐51 and IAEA TRS‐398 recommend the use of two formulas developed by Burns et al [Med. Phys. 23, 489–501 (1996)] to estimate the water‐to‐air stopping‐power ratios (SPRs). Both formulas are based on a fit to SPRs calculated for standard electron beams. This study aims to find: (1) if the formulas are applicable to beams used in TSET and (2) the impact of the ICRU report 90 recommendations on the SPRs for these beams.Methods: The EGSnrc Monte Carlo code system is used to generate 6 MeV high dose rate total skin electron (HDTSe) beams used in TSET. The simulated beams are used to calculate dose distributions and SPRs as a function of depth in a water phantom. The fitted SPRs using the empirical formulas are compared with MC‐calculated SPRs.Results: The electron beam quality specifier, the depth in water at which the absorbed dose falls to 50% of its maximum value, R50, decreases approximately 1 mm for each additional 100‐cm extended distance ranging from 2.24 cm at SSD = 100 to 1.72 cm at SSD = 700 cm. For beams passing through a scatter plate, R50 is 1.76 cm (1.14) at SSD = 300 and 1.48 cm (0.85 cm) at SSD = 600 cm with an Acrylic plate thickness of 3 mm (9 mm), respectively. The discrepancy between fitted and MC‐calculated SPRs at dref as a function of R50 is 50 is within 1% at depths 50 for beams with R50 ≥ 1.14 cm. The ICRU‐90 recommendations decrease SPRs by 0.3%–0.4% compared to the use of data recommended in ICRU‐37.Conclusion: The formulas used by the major protocols are accurate enough for clinical beams used in TSET and the error caused using the formulas is 50 for depths 50 for beams used in TSET with R50 ≥ 1.14 cm. The impact of the ICRU‐90 recommendations shows a decrease of SPRs by a fraction of a percent for beams used in TSET.

This publication has 37 references indexed in Scilit: