Monte Carlo calculations of correction factors for plane‐parallel ionization chambers in clinical electron dosimetry
Author:
- 13 August 2008
- journal article
- Published by Wiley in Medical Physics
- Vol. 35 (9) , 4033-4040
Abstract
Recent standard dosimetry protocols recommend that plane‐parallel ionization chambers be used in the measurements of depth‐dose distributions or the calibration of low‐energy electron beams with beam quality . In electron dosimetry protocols with the plane‐parallel chambers, the wall correction factor, , in water is assumed to be unity and the replacement correction factor, , is taken to be unity for well‐guarded plane‐parallel chambers, at all measurement depths. This study calculated and for NACP‐02, Markus, and Roos plane‐parallel chambers in clinical electron dosimetry using the EGSnrc Monte Carlo code system. The values for the plane‐parallel chambers increased rapidly as a function of depth in water, especially at lower energy. The value around for NACP‐02 was about 10% greater than unity at . The effect was smaller for higher electron energies. Similarly, values with depth increased drastically at the region with the steep dose gradient for lower energy. For Markus departed more than 10% from unity close to due to the narrow guard ring width. for NACP‐02 and Roos was close to unity in the plateau region of depth‐dose curves that includes a reference depth, . It was also found that the ratio of the dose to water and the dose to the sensitive volume in the air cavity for the plane‐parallel chambers, , at differs significantly from that assumed by electron dosimetry protocols.Keywords
Funding Information
- Grant in-Aid for Scientific Research (C) (19591460)
This publication has 16 references indexed in Scilit:
- Calculation of the replacement correction factors for ion chambers in megavoltage beams by Monte Carlo simulationMedical Physics, 2008
- Validation of a Monte Carlo model of a NACP-02 plane-parallel ionization chamber model using electron backscatter experimentsPhysics in Medicine & Biology, 2008
- Monte Carlo calculations of beam quality correction factorskQfor electron dosimetry with a parallel-plate Roos chamberPhysics in Medicine & Biology, 2008
- Monte Carlo study of correction factors for the use of plastic phantoms in clinical electron dosimetryMedical Physics, 2007
- An empirical method for the determination of wall perturbation factors for parallel-plate chambers in high-energy electron beamsPhysics in Medicine & Biology, 2006
- Perturbation correction factors for the NACP-02 plane-parallel ionization chamber in water in high-energy electron beamsPhysics in Medicine & Biology, 2006
- The IPEM code of practice for electron dosimetry for radiotherapy beams of initial energy from 4 to 25 MeV based on an absorbed dose to water calibrationPhysics in Medicine & Biology, 2003
- The effect of scattering foil parameters on electron‐beam Monte Carlo calculationsMedical Physics, 2001
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- Perturbation correction factors for some ionization chambers commonly applied in electron beamsPhysics in Medicine & Biology, 1991