Cochlear Implantation in the Presence of a Programmable Ventriculoperitoneal Shunt
Authors:
- Hussam K. El-KashlanHEHussam K. El-Kashlan
- 1 September 2009
- journal article
- case report
- Published by Wolters Kluwer Health in Otology & Neurotology
- Vol. 30 (6) , 704-707
Abstract
Document challenges of cochlear implantation in the presence of a ventriculoperitoneal shunt (VPS) with a programmable valve. A 7-month-old, postmeningitic, hydrocephalic infant with profound sensorineural hearing loss. Ventriculoperitoneal shunt revision with a nonprogrammable valve and bilateral cochlear implantation. Magnetic fields generated by cochlear implant (CI) transmitting coil magnets were measured ex vivo. Successful implantation and elimination of potential hazardous interaction between CI and VPS. Index patient reviewed and magnetic field strength presented. Strong magnetic fields can potentially alter the pressure setting of programmable shunt valves, causing overdrainage or underdrainage and potential serious consequences. Clear communication between the involved medical specialties is required for optimal management of patients requiring both VPS and CI.Keywords
This publication has 4 references indexed in Scilit:
- Clinical Experience in the Treatment of Idiopathic Normal-pressure Hydrocephalus Using the Programmable Gravity-assisted Valve (ProGAVAesculap)Neurosurgery Quarterly, 2007
- Programmable shunt-related suicide attemptActa Neurochirurgica, 2006
- Alteration of the Pressure Setting of a Codman-Hakim Programmable Valve by a Television-Case Report-Neurologia medico-chirurgica, 2006
- Electromagnetic field hazards involving adjustable shunt valves in hydrocephalusJournal of Neurosurgery, 2002