Subcutaneous and continuous blood pressure monitoring in an ambulatory sheep by piezoelectric micromachined ultrasonic transducers
Open Access
- 6 November 2025
- journal article
- research article
- Published by Springer Nature in Microsystems & Nanoengineering
- Vol. 11 (1) , 1-12
Abstract
This paper presents subcutaneous and continuous blood pressure (BP) monitoring using aluminum nitride (AlN) piezoelectric micromachined ultrasonic transducers (PMUTs) in an ambulatory sheep. A 37 $$\times$$ 45 PMUTs array with a footprint of 5 $$\times$$ 5 mm 2 has been designed and fabricated as a prototype device. The deep reactive ion etching (DRIE) process to open the backside holes on the silicon substrate has been optimized to create active device diaphragms with a radius of 29 μm. The resulting PMUT unit has a measured resonant frequency of 6.5 MHz in water, an output acoustic pressure of 28 kPa at a distance of 10 mm, and a 6-dB bandwidth of about 33%. The BP monitoring scheme is validated through both in vitro and in vivo experiments to illustrate the correlation between the diameter of the blood vessel and pressure. Simulations indicate that possible issues in misalignment between the device and the blood vessel can result in a 60% reduction in signal strength with only 1 mm in misalignment. This highlights the advantage of subcutaneous implantation in maintaining a stable interface and consistent alignment for reliable long-term BP monitoring, in contrast to similar approaches via wearable system setups. The in vivo testing result shows BP wave fine features such as dicrotic notches and the averaged systolic/diastolic pressure errors are −1.2 $$\pm$$ 2.1 and −2.9 $$\pm$$ 1.4 mmHg, respectively, which meets the clinical standard as calibrated by a gold-standard arterial line pressure sensor. As such, this system highlights the potential applications in silent, continuous, and highly accurate BP monitoring for hypertension patients using this implantable MEMS-based technology.
Keywords
Funding Information
- Berkeley Sensor and Actuator Center
This publication has 55 references indexed in Scilit:
- Tissue‐Adhesive Piezoelectric Soft Sensor for In Vivo Blood Pressure Monitoring During Surgical OperationAdvanced Functional Materials, 2023
- Assessment of Long-term Benefit of Intensive Blood Pressure Control on Residual Life SpanJAMA Cardiology, 2020
- PDMS microfluidics: A mini reviewJournal of Applied Polymer Science, 2020
- A review on coronary artery disease, its risk factors, and therapeuticsJournal of Cellular Physiology, 2019
- A perspective on molybdenum biocompatibility and antimicrobial activity for applications in implantsJournal of Materials Science, 2015
- Wireless Blood Pressure Monitoring with a Novel Implantable Device: Long-Term In Vivo ResultsCardioVascular and Interventional Radiology, 2014
- Continuous in vivo blood pressure measurements using a fully implantable wireless SAW sensorBiomedical Microdevices, 2013
- Relationship Between Measurement Site and Motion Artifacts in Wearable Reflected PhotoplethysmographyJournal of Medical Systems, 2010
- Transducers and Arrays for Underwater Sound2007
- The diameter of the intact carotid artery in man and its change with pulse pressurePflügers Archiv - European Journal of Physiology, 1968