Quantification of near‐field heating during volumetric MR‐HIFU ablation
Authors:
- C. MougenotCMC. Mougenot
- M. O. KöhlerMKM. O. Köhler
- J. EnholmJEJ. Enholm
- B. QuessonBQB. Quesson
- C. MoonenCMC. Moonen
- 20 December 2010
- journal article
- Published by Wiley in Medical Physics
- Vol. 38 (1) , 272-282
Abstract
Purpose: High‐intensity focused ultrasound guided by magnetic resonance imaging has been extensively evaluated during the past decade as a clinical alternative for thermal ablation of tumor tissue. However, the maximal ablation volume is limited by the extensive treatment duration resulting from the small size of the focal point as compared to the average tumor size. Volumetric sonication has been shown to efficiently enlarge the ablated volume per sonication, but remains limited by the temperature increase induced in the skin and fat layers. In this study, multiplane MR thermometry is proposed for monitoring the near‐field temperature rise in order to prevent related unintended thermal damage.Methods: The method was evaluated by performing sonications in the thigh muscle of 11 pigs maintained under general anesthesia. Volumetric ablations were performed by steering the focal point along trajectories consisting of multiple outward‐moving concentric circles. Near‐field heating was characterized with MR temperature maps and thermal dose maps. The results from the MR measurements were compared to simulations.Results: In this study, the measured maximum temperature rise was found to correlate linearly with the surface energy density within the near field of the beam path with a slope of. This simple linear model appears to be almost independent of the trajectory pattern and the sonication depth. The safety limit to avoid lethal damage of the subcutaneous tissues of the porcine thigh was identified to be an absolute temperature of , corresponding to a surface energy density of at 1.2 MHz.Conclusions: A linear relationship can be established to estimate the temperature increase based on the chosen power prior to ablation, thereby providing ana priori safety check for possible excessive near‐field heating using a known surface energy density threshold. This method would also give the clinician the possibility to abort the sonication should excessive near‐field temperature rise be seen before fat layer damage or skin burns are inflicted.Keywords
This publication has 27 references indexed in Scilit:
- Feasibility of magnetic resonance imaging-guided high intensity focused ultrasound therapy for ablating uterine fibroids in patients with bowel lies anterior to uterusEuropean Journal of Radiology, 2010
- Non‐invasive determination of tissue thermal parameters from high intensity focused ultrasound treatment monitored by volumetric MRI thermometryNMR in Biomedicine, 2009
- Monitoring and correcting spatio-temporal variations of the MR scanner’s static magnetic fieldMagnetic Resonance Materials in Physics, Biology and Medicine, 2006
- Extracorporeally induced ablation of renal tissue by high‐intensity focused ultrasoundBJU International, 2006
- Determination of heterogeneous thermal parameters using ultrasound induced heating and MR thermal mappingPhysics in Medicine & Biology, 2006
- Full thickness abdominal burn following magnetic resonance guided focused ultrasound therapyBurns, 2005
- A precise and fast temperature mapping using water proton chemical shiftMagnetic Resonance in Medicine, 1995
- Compilation of empirical ultrasonic properties of mammalian tissues. IIThe Journal of the Acoustical Society of America, 1980
- Comprehensive compilation of empirical ultrasonic properties of mammalian tissuesThe Journal of the Acoustical Society of America, 1978
- Theory of Focusing RadiatorsThe Journal of the Acoustical Society of America, 1949