Optimization of white-matter-nulled magnetization prepared rapid gradient echo (MP-RAGE) imaging
Open Access
- 29 May 2014
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 73 (5) , 1786-1794
Abstract
Purpose To optimize the white‐matter‐nulled (WMn) Magnetization Prepared Rapid Gradient Echo (MP‐RAGE) sequence at 7 Tesla (T), with comparisons to 3T. Methods Optimal parameters for maximizing signal‐to‐noise ratio (SNR) efficiency were derived. The effect of flip angle and repetition time (TR) on image blurring was modeled using simulations and validated in vivo. A novel two‐dimensional (2D) ‐centric radial fan beam (RFB) k‐space segmentation scheme was used to shorten scan times and improve parallel imaging. Healthy subjects as well as patients with multiple sclerosis and tremor were scanned using the optimized protocols. Results Inversion repetition times (TS) of 4.5 s and 6 s were found to yield the highest SNR efficiency for WMn MP‐RAGE at 3T and 7T, respectively. Blurring was more sensitive to flip in WMn than in CSFn MP‐RAGE and relatively insensitive to TR for both regimes. The 2D RFB scheme had 19% and 47% higher thalamic SNR and SNR efficiency than the 1D centric scheme for WMn MP‐RAGE. Compared with 3T, SNR and SNR efficiency were higher for the 7T WMn regime by 56% and 41%, respectively. MS lesions in the cortex and thalamus as well as thalamic subnuclei in tremor patients were clearly delineated using WMn MP‐RAGE. Conclusion Optimization and new view ordering enabled MP‐RAGE imaging with 0.8–1 mm3 isotropic spatial resolution in scan times of 5 min with whole brain coverage. Magn Reson Med 73:1786–1794, 2015.Keywords
Funding Information
- NIH (NIH P41 EB015891-19)
- Richard M. LUCAS Foundation
- GE Healthcare
This publication has 19 references indexed in Scilit:
- Automated quantification of white matter disease extent at 3 T: Comparison with volumetric readingsJournal of Magnetic Resonance Imaging, 2012
- 3 Tesla and 7 Tesla MRI of multiple sclerosis cortical lesionsJournal of Magnetic Resonance Imaging, 2010
- The Alzheimer's disease neuroimaging initiative (ADNI): MRI methodsJournal of Magnetic Resonance Imaging, 2008
- Effect of windowing and zero‐filled reconstruction of MRI data on spatial resolution and acquisition strategyJournal of Magnetic Resonance Imaging, 2001
- Optimization of 3-D MP-RAGE Sequences for Structural Brain ImagingNeuroImage, 2000
- Quantitative volumetric analysis of brain magnetic resonance imaging from patients with multiple sclerosisJournal of the Neurological Sciences, 1998
- Adiabatic pulsesNMR in Biomedicine, 1997
- Multiarchitectonic and stereotactic atlas of the human thalamusJournal of Comparative Neurology, 1997
- The importance of phase‐encoding order in ultra‐short TR snapshot MR imagingMagnetic Resonance in Medicine, 1990
- Three‐dimensional magnetization‐prepared rapid gradient‐echo imaging (3D MP RAGE)Magnetic Resonance in Medicine, 1990