The analysis of particles at low accelerating voltages (<= 10 kV) with energy dispersive x-ray spectroscopy (EDS)
Author:
- 1 November 2002
- journal article
- Published by National Institute of Standards and Technology (NIST) in Journal of Research of the National Institute of Standards and Technology
- Vol. 107 (6) , 555-66
Abstract
In recent years, there have been a series of advancements in electron beam instruments and x-ray detectors which may make it possible to improve significantly the quality of results from the quantitative electron-probe analysis of individual particles. These advances include: (1) field-emission gun electron beam instruments such as scanning electron microscopes (FEG-SEMs) that have high brightness electron guns with excellent performance at low beam energies, E 0 ≤ 10 keV and (2) high-resolution energy-dispersive x-ray spectrometers, like the microcalorimeter detector, that provide high-resolution (< 10 eV) parallel x-ray collection. These devices make it possible to separate low energy (< 4 keV) x-ray lines including the K lines of carbon, nitrogen and oxygen and the L and M lines for elements with atomic numbers in the range of 25 to 83. In light of these advances, this paper investigates the possibility of using accelerating voltages ≤ 10 kV, as a method to improve the accuracy of elemental analysis for micrometer-sized particles.Keywords
This publication has 2 references indexed in Scilit:
- Preparation and Certification of K-411 Glass MicrospheresMicroscopy and Microanalysis, 2000
- Measures for spectral quality in low‐voltage X‐ray microanalysisScanning, 2000