How Does the Plethysmogram Derived from the Pulse Oximeter Relate to Arterial Blood Pressure in Coronary Artery Bypass Graft Patients?
Authors:
- M. Ashraf M. GhobashyMGM. Ashraf M. Ghobashy
- Robert G. StoutRSRobert G. Stout
- David G. SilvermanDSDavid G. Silverman
- Kirk H. ShelleyKSKirk H. Shelley
- 1 December 2001
- journal article
- research article
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 93 (6) , 1466-1471
Abstract
Twenty patients scheduled for coronary artery bypass grafting had their ear and finger oximeter and radial artery blood pressure (Bpmeas) waveforms collected. The ear and finger pulse oximeter waveforms were analyzed to extract beat-to-beat amplitude and area and width measurements. The Bpmeas waveforms were analyzed to measured systolic blood pressure (BP), mean BP, and pulse pressure. The correlation coefficient was determined between the derived waveforms from the pulse oximeter and Bpmeas for the first 10 patients. The ear pulse oximeter width (WidthEar) had the best correlation (r = 0.8). Linear regression was done between WidthEar and Bpmeas based on slope (b) and intercept (a) values, BP was calculated (Bpcalc) in the next 10 patients as:MATHwhere i = systolic BP, mean BP, and pulse pressure. The initial bias was too large to be clinically useful. To improve clinical applicability a period of calibration was introduced in which the first 50 readings of WidthEar and Bpmeas for each patient were used to calculate the intercept. After calibration the systolic BP, mean BP and pulse pressure bias values were −2.6, −1.88 and −1.28 mm Hg, and the precision values were 15.9 10.09, and 9.94 mm Hg, respectively. The present attempt to develop a clinically useful method of noninvasive BP measuring was partly successful with the requirement of a calibration period.Keywords
This publication has 20 references indexed in Scilit:
- Accuracy and precision of blood pressure determination with the Finapres: an overview using re-sampling statisticsJournal of Human Hypertension, 1998
- Autonomic Control of Skin Microvessels: Assessment by Power Spectrum of Photoplethysmographic WavesClinical Science, 1996
- Arterial Tonometry for Noninvasive, Continuous Blood Pressure Monitoring during AnesthesiaAnesthesiology, 1991
- The Accuracy of Finapres™ Noninvasive Mean Arterial Pressure Measurements in Anesthetized PatientsAnesthesiology, 1991
- Comparing Methods of Measurement: An Alternative ApproachAnesthesiology, 1990
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Evaluation of Pulse OximetryAnesthesiology, 1983
- Measurement in Medicine: The Analysis of Method Comparison StudiesJournal of the Royal Statistical Society: Series D (The Statistician), 1983
- On the Safety of Radial Artery CannulationAnesthesiology, 1983
- Spectrophotometric monitoring of arterial oxygen saturation in the fingertipMedical & Biological Engineering & Computing, 1980