Retinal oximeter for the blue-green oximetry technique
Open Access
- 1 January 2011
- journal article
- research article
- Published by SPIE-Intl Soc Optical Eng in Journal of Biomedical Optics
- Vol. 16 (10) , 107004-107004-7
Abstract
Retinal oximetry offers potential for noninvasive assessment of central venous oxyhemoglobin saturation (SO2) via the retinal vessels but requires a calibrated accuracy of ±3% saturation in order to be clinically useful. Prior oximeter designs have been hampered by poor saturation calibration accuracy. We demonstrate that the blue-green oximetry (BGO) technique can provide accuracy within ±3% in swine when multiply scattered light from blood within a retinal vessel is isolated. A noninvasive on-axis scanning retinal oximeter (ROx-3) is constructed that generates a multiwavelength image in the range required for BGO. A field stop in the detection pathway is used in conjunction with an anticonfocal bisecting wire to remove specular vessel reflections and isolate multiply backscattered light from the blood column within a retinal vessel. This design is tested on an enucleated swine eye vessel and a retinal vein in a human volunteer with retinal SO2 measurements of ∼1 and ∼65%, respectively. These saturations, calculated using the calibration line from earlier work, are internally consistent with a standard error of the mean of ±2% SO2. The absolute measures are well within the expected saturation range for the site (–1 and 63%). This is the first demonstration of noninvasive on-axis BGO retinal oximetry.Keywords
This publication has 38 references indexed in Scilit:
- Glaucoma Filtration Surgery and Retinal Oxygen SaturationInvestigative Ophthalmology & Visual Science, 2009
- Blue-green spectral minimum correlates with oxyhemoglobin saturation in vivoJournal of Biomedical Optics, 2008
- Diffuse spectral fundus reflectance measured using subretinally placed spectralonJournal of Biomedical Optics, 2008
- Blood oxyhemoglobin saturation measurements by blue-green spectral shiftJournal of Biomedical Optics, 2007
- Automatic Retinal OximetryInvestigative Ophthalmology & Visual Science, 2006
- Intravascular oxygen saturation in retinal vessels in normal subjects and open‐angle glaucoma subjectsActa Ophthalmologica Scandinavica, 2006
- Retinal Oximetry Using Intravitreal IlluminationCurrent Eye Research, 2006
- In Vitro Multispectral Diffuse Reflectance Measurements of the Porcine FundusInvestigative Ophthalmology & Visual Science, 2005
- Retinal Venous Oxygen Saturation Correlates with Blood VolumeAcademic Emergency Medicine, 1998
- A Study of Retinal Venous Blood Oxygen Saturation in Human Subjects by Photographic MeansCirculation, 1963